> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000002421.md).

# Thawing, stability, and storage of NovaSeq 6000 reagents

For information regarding the thawing, stability, and storage of NovaSeq 6000 reagents, see below.

Reagents in the SBS and buffer cartridges are sensitive to light. Protect those cartridges from light by storing them in a dark location away from any light sources.

**Temperature key**

| Storage Condition     | Temperature range °C |
| --------------------- | -------------------- |
| Room temperature (RT) | 15°C to 30°C         |
| Refrigerator (4°C)    | 2°C to 8°C           |
| Freezer (-20°C)       | -25°C to -15°C       |

**Kit Components, shipping conditions, and storage temperatures**

| Quantity | Kit Component     | Shipping Condition | Storage Temp   | Notes                                                                                                                                                                                       |
| -------- | ----------------- | ------------------ | -------------- | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
| 1        | Library tube      | Room temp          | 15°C to 30°C   |                                                                                                                                                                                             |
| 1        | Buffer cartridge  | Room temp          | 15°C to 30°C   | Light sensitive.                                                                                                                                                                            |
| 1        | Flow cell         | Ice pack           | 2°C to 8°C     | <p>Before use, allow the flow cell to sit at room temperature for 10-15 minutes before removing from the packaging.<br><br>Once open, use within 12 hours of removing from the package.</p> |
| 1        | Cluster cartridge | Dry ice            | -25°C to -15°C |                                                                                                                                                                                             |
| 1        | SBS Cartridge     | Dry ice            | -25°C to -15°C | Light sensitive.                                                                                                                                                                            |

**Thawing guidelines for each kit component**

For the components that are thawed in a water bath, place them in a room temperature water bath (19°C - 25°C) following the timing guidance specified below.

**Cluster cartridges** \*Shipped with dry ice (store at -25°C to -15°C).

| Cluster cartridge type | Duration of thaw                                                   |
| ---------------------- | ------------------------------------------------------------------ |
| SP, S1, S2             | Up to 2 hours in a room temperature (RT) water bath (19°C - 25°C). |
| S4                     | Up to 4 hours in room temperature (RT) water bath (19°C - 25°C).   |

* Use a wire rack to prevent tipping.
* If reagents cannot be loaded onto the instrument within 4 hours, store at 2°C to 8°C for up to 24 hours.
* Can refreeze once, if needed. If refreezing, do so immediately after thawing.

**SBS cartridges** \*Shipped with dry ice (store at -25°C to -15°C).

| SBS cartridge type | Duration of thaw                                          |
| ------------------ | --------------------------------------------------------- |
| SP, S1, S2         | 4 hours in room temperature (RT) water bath (19°C - 25°C) |
| S4                 | 4 hours in room temperature (RT) water bath (19°C - 25°C) |

* Use a wire rack to prevent tipping.
* If reagents cannot be loaded onto the instrument within 4 hours, store at 2°C to 8°C for up to 24 hours.
* Can refreeze once, if needed. If refreezing, do so immediately after thawing.
* SBS cartridges are sensitive to light.

**Buffer Cartridge** \*Shipped at RT (store at RT).

* Buffer cartridges are sensitive to light.
* Buffer cartridges can be recycled (only consumable that can be recycled).

**Flow cell (FC)** \*Shipped dry with an ice pack (store at 2-8°C).

* Before use, allow the flow cell to sit at room temperature for 10-15 minutes before removing it from the packaging.
* Once open, use within 12 hours of removing it from the package.

**Library tube** \*Shipped at RT (store at RT).

* Store the library tube (containing the library pool) only if necessary.
* Long-term library pool storage at -25°C to -15°C can increase duplicates, which decreases yield.

For further information, see Knowledge Base articles [Xp reagents components and stability on the NovaSeq 6000](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000002422.md) and [How to thaw and store sequencing reagents for optimal performance.](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001533.md)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2421), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002421%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000002421.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.