> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000007819.md).

# How to troubleshoot CIB Fault Initialization errors on NovaSeq X Series instruments

The NovaSeq X Series Control Software will verify a stable connection to the Camera Interface Board (CIB) during initialization. This results in one of three outcomes:

1. The CIB responds as expected, and initialization will proceed normally.
2. The CIB times out or is in a bad state, but the Software automatically recovers the connection.
3. The CIB times out and the Software is not able to recover the connection.

**If the CIB Fault is recovered automatically**

If the CIB times out or is in a bad state, the NovaSeq X Series Control Software will automatically attempt to recover the CIB connection. If this recovery attempt is successful, the following error will be displayed.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-9a6d5aba604eaf0e69b323eadd51139bf3a83d3a%2Fimage1.png?alt=media\&token=f4816037-c016-401c-bbff-3f1b335ff4d4)

If this error presents, see the following Troubleshooting Actions:

1. **Do not** shutdown or power cycle the instrument.
2. Select **Exit Application** and restart the NovaSeq X Series Control Software.
3. If initialization is successful, proceed with normal instrument workflow.

**If the CIB connection does not recover automatically**

If the Control Software is not able to recover the CIB connection, it will interrupt software initialization with the following error.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-fba12f3761b7f656b63f16825c6c79770827f76a%2Fimage2.png?alt=media\&token=8526e12e-c187-4ace-b45b-88bd6eaec97f)

If this error presents, see the following Troubleshooting Actions:

1. **Do not** shutdown or power cycle the instrument.
2. Select **Exit Application** and restart the NovaSeq X Series Control Software.
   1. If initialization is successful, proceed with normal instrument workflow.
   2. If the error persists, repeat Step 2 up to two additional times.
3. If the issue persists after three attempts to restart the NovaSeq X Series Control Software, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7819), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007819%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000007819.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.