> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-reference_material-list/000002063.md).

# Recommendations for using custom primers on the NextSeq 500/550

This article provides an overview of Custom Primer use for the NextSeq500/550. For complete Custom Primer guidance, please review the [NextSeq 500 and NextSeq 550 Sequencing Systems Custom Primers Guide](https://support.illumina.com/downloads/nextseq-500-custom-primers-guide-15057456.html).

* Custom primer positions in the reagent cartridge:
  * Custom Read 1 primer: position 7
  * Custom Read 2 primer: position 8
  * Custom Index 1/2 primers: position 9
* Can only one custom index primer be used?
  * With NextSeq Control Software (NCS) v4.0+, one or both custom index primers can be used. For example, the following setups are possible:
    * Custom i7 only
    * Custom i5 only
    * Both custom i7 and custom i5
* Are there rules to using custom primers on the NextSeq 500/550?

  * Rules will be dependent upon the construction of the library.
  * For example, using a custom Read 1 primer assumes a custom P5 adapter, which might also require a custom i5 index primer. However, depending on the library design, this might not be true.

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-b48eb97774d820b93acfea40a4989b7f7e4f0b3c%2Fimage1.png?alt=media\&token=10e996a3-7812-44fa-801e-09bbaab5fd00)

Note: Using a custom primer for Index 2 Read is not supported with other combinations or NCS versions earlier than v1.4.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2063), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002063%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-reference_material-list/000002063.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.