> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000002112.md).

# Troubleshooting a flow cell registration stall during pre-run checks on first run on NextSeq 500/550

On the first run after upgrading the NextSeq 500/550 to NextSeq Control Software (NCS) v4.0.X, the pre-run checks can take around 30 minutes to complete flow cell registration. This is expected, as the instrument will be downloading drives to perform the check.

Allow the pre-run check to continue as long as no errors appear.

If an option appears to retry the pre-run check, this may indicate a communication issue. This can be resolved by reseating the flow cell and power cycling the instrument.

For further information, see [How to Power Cycle the NextSeq 500/550](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001972.md).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2112), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002112%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000002112.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.