# Troubleshooting pre run checks with Temperature sensor check failed: Board 0/1 on NextSeq 500/550

In rare instances, it is possible for the NextSeq 500/550 to fail to successfully complete the pre-run check with a "Temperature sensor check failed: Originally caused by: Temperature is out of range for Board 0 and/or Board 1 Temperature Probe" error message.

The Main Board (MPCA) assists with connecting the onboard computer with the instrument hardware and includes two temperature sensors integrated into the MPCA. This error may display as a result of a communication error, or potentially a defective sensor.

![](/files/XOzAyR2b3GLgDwHU4Cb1)

**Troubleshooting steps:**

1. Power cycle the NextSeq 500/550.

* Search Knowledge Base Article **How to Perform a NextSeq 500/550 Power Cycle**.

2. Repeat the run setup.
3. If error returns after a power cycle, or returns on subsequent run setup, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for additional assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #6373), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000006373%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000006373.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.