> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/rna-library-prep/library-preparation-rna-library-prep-faq-list/000001155.md).

# TruSeq Stranded RNA bead clean-up and library size information

**What to do** **if the clean up beads are not dry after the 15 minutes incubation at room temperature?**\
Longer drying times may be required depending on environmental variables and the amount of ethanol remaining in the well. However, take care not to over-dry beads as this can impact sample recovery. Fewer than 15 minutes may be sufficient for drying.

**What is the size of adapter dimers with the dual indexed adapters?**\
The adapter dimers will be approximately 136 bp, may run closer to 150 bp on a trace.

**Is it possible to generate larger RNA fragment sizes?**\
It is possible to use larger fragment sizes. However, Illumina has found that shorter fragment sizes generate the most even transcript coverage. Refer to the Appendix section of the [TruSeq Stranded Total RNA Reference Guide](https://support.illumina.com/downloads/truseq-stranded-total-rna-reference-guide-1000000040499.html) for more information.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1155), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001155%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/rna-library-prep/library-preparation-rna-library-prep-faq-list/000001155.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.