> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000006639.md).

# Flow cell not detected error on the NextSeq 1000/2000

**Background:** When starting a run, the NextSeq 1000/2000 can present a "flow cell not detected" error despite a properly loaded flow cell.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-f3f35acd2d1b1b94a21618ab9b94c1087ca122f0%2Fimage1.jpg?alt=media)

**To troubleshoot:**

1. Investigate the flow cell seating and cleanliness.
   * Retry by reseating the flow cell.
2. Power cycle\* the system.
   * Confirm failure to recognize the flow cell persists through the power cycle.
   * \*For further information, see the Knowledge Base articles [How to power cycle the NextSeq 1000/2000](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000003038.md) and [How To Power Cycle NextSeq 1000/2000 Video](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000005552.md).
3. If issue persists, prepare the following information and contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for further guidance.

**When reaching out to Illumina Technical Support, provide the following:**

* **Control Software Logs**: /usr/local/illumina/logs/nextseq1k2k-control-software
* **Service Software Logs**: /usr/local/illumina/logs/nextseq1k2k-service-software

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #6639), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000006639%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000006639.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.