Flow cell is not detected error on the NextSeq 1000/2000
Last updated
Was this helpful?
Last updated
Was this helpful?
Background: When starting a run, the NextSeq 1000/2000 can present an error that the "flow cell is not detected" despite a properly loaded flow cell.
To troubleshoot:
Investigate the flow cell seating and cleanliness.
Retry by reseating the flow cell.
Power cycle* the system.
Confirm failure to recognize the flow cell persists through the power cycle.
If issue persists, prepare the requested information below and contact Illumina Technical Support for further guidance.
When reaching out to Illumina Technical Support, provide the following:
Control Software Logs: /usr /local/Illumina/Logs/nextseq1k2k-control-software
Service Software Logs: /usr /local/Illumina/Logs/nextseq1k2k-service-software
*For further information, see the Knowledge Base articles and.
For any feedback or questions regarding this article (Illumina Knowledge Article #6639), contact Illumina Technical Support .