> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009260.md).

# Troubleshooting Internal Server Error in Cloud Settings on the MiSeq i100 Series

**Background**

BaseSpace Sequencing Hub leverages several global data centers for regional data storage and processing. The MiSeq i100 Control Software will contact BaseSpace Ground Control to get a list of available regions and use this to populate the Hosting Location dropdown menu in the Cloud Settings page.

If there are issues with connecting to Ground Control, the Control Software will be unable to populate the Hosting Location dropdown menu and will present a Internal Server error message at the top of the Cloud Settings menu.

"An unexpected error occurred. Details: Internal Server Error."

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-19f341e4bf10e42a6ece6137644433bd2fe67fa7%2Fimage1.png?alt=media)

Please note that User Interface (UI) does not load properly when this error presents and the checkboxes will all appear to be de-selected, though this might not reflect the current Cloud Settings configuration status.

To resolve this issue, please see the following **Troubleshooting Actions**. If further assistance is required, contact [Illumina Tech Support](https://www.illumina.com/company/contact-us.html).

**Troubleshooting Actions**

1. Consult Illumina's [Product Security Portal](https://support.illumina.com/product-security-portal.html) documentation for recommended URL and port allowlists required for Proactive and BaseSpace Sequencing Hub access for the MiSeq i100
2. Contact your local IT team and request that the URLs under the appropriate regional **Hosting Location** are added to your network security allow list.
3. Once this is complete, navigate to the Cloud Settings menu in the MiSeq i100 Control Software.
4. Confirm that the Hosting Location dropdown menu is now populated and then select the appropriate regional Hosting Location.
   1. Is using a private Enterprise domain, input the private domain name in the free text field.
5. Select **Test Configuration** and confirm that all required URLs are accessible.
6. Ensure that the checkbox next to **Send instrument performance data to Illumina** is selected to enable Proactive support.
7. Ensure that the appropriate checkboxes under **Run Settings** are selected to enable Cloud Run Monitoring or Cloud Run Storage.
8. Select **Save**.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9260), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009260%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009260.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.