For the complete documentation index, see llms.txt. This page is also available as Markdown.

Troubleshooting NovaSeq 6000 reagent RFID failures

Background: NovaSeq 6000 consumables are labeled with Radio-frequency identification (RFID) tags that the instrument uses to identify them. Occasionally a run setup error can occur where the instrument is unable to read the RFID from the loaded consumable. This article describes how to troubleshoot this error.

To troubleshoot a cartridge RFID failure:

  • Load a previously used consumable.

    • If the used consumable is detected as previously used, the sensor is working as expected. In that case, load new reagent/buffer cartridge and contact Illumina Technical Support for further assistance. Provide the cartridge product number, lot number, and serial number to Illumina Technical Support.

    • If the used consumable replicates the "RFID cannot be read" error, the sensor may be offline. If the instrument is idle on the other side, perform a power cycle*. Once the instrument initializes:

      • Retry the pre-run check with the same used cartridge to check if the sensor is online.

      • If the used cartridge is now detected as "used," load the original cartridge and start the pre-run check again.

      • If the error persists after trying a used consumable and power cycling the instrument, contact Illumina Technical Support for the next steps.

*For further information, see Knowledge Base article How to Power Cycle the NovaSeq 6000.

To troubleshoot aflow cell RFID failure:

  • Inspect flow cell visually to identify any debris or glass shards on the surface and underside.

    • If debris or glass shards are present:

      • Clean the bottom of the flow cell and flow cell stage with an alcohol wipe. Pat dry excess alcohol with a lint-free wipe. Make sure only to clean the glass surface of the flow cell.

      • Inspect the flow cell mount for debris or shards. Attempt to clean this area like the flow cell, avoiding pushing any debris into the vacuum holes on the mount

      • Retry the pre-run check after performing the above steps with the same flow cell.

      • If the pre-run check fails again with the same error, try setting up the run with a new flow cell. Contact Illumina Technical Support for further assistance. Provide flow cell product number, lot number, and serial number.

    • If debris or glass shards are not present:

      • Exit out of run setup and from the control software and power cycle the instrument.

      • Launch the NovaSeq Control Software (NVCS) and retry run setup

      • If error persists with the original flow cell, attempt the cleaning procedure above and then retry run setup with the original flow cell

      • If the error persists, try a new flow cell. If the run starts with a new flow cell, contact Illumina Technical Support for further assistance.

      • If the error persists with a newflow cell, contact Illumina Technical Support for assistance.

For any feedback or questions regarding this article (Illumina Knowledge Article #1297), contact Illumina Technical Support techsupport@illumina.com.

Last updated

Was this helpful?