> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-reference_material-list/000002258.md).

# How to enable Wi-Fi on the iSeq 100

The iSeq 100 has integrated wireless networking hardware. By default, the Wi-Fi adapter is disabled because a wired connection is recommended for the most reliable data transfer. The Wi-Fi can be enabled using the following steps. Enabling wireless connection requires admin privileges, ensure user is logged in as sbsadmin or an account with admin privileges (not sbsuser).

**Note**: Wireless connectivity is intended for internet access and post-run data transfer. It is not recommend to select an output folder via wireless connection.

1. From the iSeq Control Software, click or tap on Menu on the top left and select **System Settings**

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-32edb6f29c76860dba1f8a070ac8ab3c9598117a%2Fimage1.png?alt=media)

2. Then select the **Network Access** tab from the top of the menu
3. Select the **Network Configuration** button

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-457328a37b1231946ea32ec6c76a990ea205a811%2Fimage2.png?alt=media)

4. On the Windows network & internet status screen, select **Change adapter options**

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-f381c236c70b07ae023479839c4475fbd22d0763%2Fimage3.png?alt=media)

5. Enable the Wi-Fi adapter by right clicking or long pressing on it and selecting **Enable**

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-2c5ed8748e35459cbea9580115985ebcedbcd4fe%2Fimage4.png?alt=media)

6. Close network connections window after Wi-Fi adapter is enabled and navigate back to Windows Network & Internet
7. Configure Wi-Fi as needed from within the Wi-Fi tab

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-af4639aa0f7f109f0e1fa2dbcf2a09f068e8e096%2Fimage5.png?alt=media)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2258), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002258%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-reference_material-list/000002258.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.