> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001324.md).

# How to requeue and import run data for reanalysis with Local Run Manager

The Local Run Manager software is an integrated solution for recording samples for a run, specifying run parameters, monitoring run status, performing data analysis, and viewing results. This article describes how to requeue analysis for a run in Local Run Manager with the same module, and how to import data to analyze with a new module.

**To requeue with the same module:**

1. From the Local Run Manager dashboard, select **Actions** next to the run you wish to requeue. Select **Requeue**. ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-1d86bf908f572bf79ce71bc2e1926307cfd26078%2Fimage1.jpg?alt=media)
2. In the pop-up window, choose whether to **Edit Setup** or to **Requeue** with the original setup information. ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-206718e2b663cab00def5d266049dc0578655f28%2Fimage2.png?alt=media)

* If **Edit Setup** was chosen, change the desired parameters, then select **Requeue Analysis**. The status shows “Analysis Completed” when the requeued analysis has completed

**To import data to a new analysis module:**

Importing run data can be used to reanalyze an existing run in a new workflow, or for off-instrument analysis.

1. From the dashboard, select **Create Run** and the desired module. On the next page, set the run parameters, then select **Save Run**. The run parameters must be the same as the original run. Important information to include is the read type and number of read cycles, the number and length of index reads, and whether custom primers were used. The original run setup can be determined in different ways depending on the source data:
   1. For runs previously analyzed in Local Run Manager, select the original run in the Dashboard. Select the **Samples & Results** tab then **Requeue Analysis**. Select **Edit Setup** to view the run configuration.
   2. When using Local Run Manager, the **Import Sample Sheet** option can be used to autopopulate the read and sample information.
   3. The number of read cycles for a run can be found in the RunInfo.xml file in the run folder.
   4. For more information on setting module-specific parameters, refer to the [analysis module workflow guides](https://support.illumina.com/sequencing/sequencing_software/local-run-manager/documentation.html).\
      ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-7a7312466a25ca629c1f9e3ab8f8b932d268510c%2Fimage3.png?alt=media)
2. In the dashboard, identify the run that was just created.
3. Select More Options (three vertical dots) next to the newly created run, and select **Import**. ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-75175ad73bc219c3ce0494ca71f3ad608a62e909%2Fimage4.jpg?alt=media)
4. Enter the full run folder path and the output folder location.\
   **Note:** Local Run Manager does not support mapped drives or directories, but does support Universal Naming Convention (UNC) paths. Local Run Manager may fail to import runs from locations under user-specific folders, such as Desktop and Documents. Importing runs from universally accessible locations is recommended. Runs imported from network locations may take longer to analyze than runs imported from a local drive.\
   ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-e9cef003844dfcc881cfa5b7a9e96e5638c8cfae%2Fimage5.png?alt=media)
5. Select **Import Run** to start the analysis. The status shows “Analysis Completed” when the requeued analysis has finished.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1324), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001324%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001324.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.