How to Power Cycle the NovaSeq X Series Instruments
Background
Restarting or power cycling the NovaSeq X Series instruments can resolve many common communication or instrumentation issues. However, these actions should only be performed when recommended by Illumina Tech Support or Field Service. Power cycling the instrument is expected to take up to 45 minutes.
If the NovaSeq X Series Control Software is not accessible or not responding, do not shutdown through the Linux Operating System (OS). Contact Illumina Tech Support for further guidance.
Power Cycling the NovaSeq X Series Instruments
For Control Software v1.3 and above:
IMPORTANT NOTE: For instruments running Control Software v1.3 and above, the guidance to power cycle every 14 days with a Maintenance Wash is no longer mandatory. However, it is still recommended to routinely power cycle once a month.
See the following steps to Power Cycle the NovaSeq X Series instrument if recommended by Illumina Support teams as part of normal Troubleshooting:
From the NovaSeq X Series Control Software Home screen, select the instrument icon to open the global navigation menu, and then select Settings.
From the Settings menu, select Restart Instrument.
The Control Software will close the application initiate the NovaSeq X Series instrument shutdown and then restart automatically.
DO NOT toggle the power switch on the back of the instrument, the instrument will fully power down on its own.
Once the NovaSeq X Series instrument has restarted, login to the Linux Operating System (OS)
Once logged into the Linux OS, select Activities in the top-left corner, then select the NovaSeq X Series Control Software icon to launch the Control Software.
For Control Software v1.2.2 only:
For instruments running Control Software v1.2.2 only, perform the steps listed below every 14 days with the regularly scheduled Maintenance Wash.
From the NovaSeq X Control Software Home screen, select the instrument icon to open the global navigation menu, and then select Settings.
From the Settings menu, select Shutdown.
DO NOT toggle the power switch on the back of the instrument, the instrument will fully power down on its own.
Wait until the power button on the right side of the instrument begins to pulse, then press the power button.
This step can take up to 25 minutes.
Allow the instrument to boot, then login to the Linux OS.
Once logged into the Linux OS, select Activities in the top-left corner, then select the NovaSeq X Series Control Software icon to launch the Control Software.
Instrument boot and initialization can take 20-25 minutes in total.
For Control Software v1.2 and below:
From the NovaSeq X Control Software Home screen, select the instrument icon to open the global navigation menu, and then select Settings.
From the Settings menu, select Shutdown.
Wait at least 10 minutes for the instrument to completely shutdown. The monitor should be dark, with no text or cursors flashing in the top-left corner.
When the instrument computer has powered down and the monitor is dark, toggle the power switch on the back right side of the instrument to the OFF position.

Allow the instrument to remain powered down for 30 seconds.
Toggle the power switch on the back of the instrument to the ON position.
Wait until the power button on the right side of the instrument begins to pulse, then press the power button.
Allow the instrument to boot, then login to the Linux Operating System (OS).
Allow 5 minutes for the instrument Compute Engine (CE) and Kubernetes microservices to initialize before attempting to start the Control Software.
Within the Linux OS, select Activities in the top left corner, then select the NovaSeq X Series Control Software icon and allow control software to initialize.
The Windows key on the keyboard can be used to open the Activities menu and select the NovaSeq X Series Control Software icon.
If the NovaSeq X Series Control Software does not initialize as expected after power cycling the instrument, contact Illumina Tech Support for further guidance.
For any feedback or questions regarding this article (Illumina Knowledge Article #7478), contact Illumina Technical Support techsupport@illumina.com.