> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001914.md).

# How to troubleshoot RFID errors on the NextSeq 500/550

These types of errors can occur during run setup when consumables are loaded into the instrument. These can be due to a transient communication issue with the Radio-frequency identification (RFID) sensor, a broken sensor, or a fault in the consumable RFID.

Unlike the MiSeq, there is no bypass code that can be used as an option to bypass the error on the NextSeq 500/550.

\*\*Error message:\*\*RFID tag appears to be blank: signed data layout version = 0\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-286d64cb097ec32015dbe8d1837179ed6f2386d7%2Fimage1.png?alt=media\&token=159a2534-7135-48f6-801e-a7e0cd23c647)

The error may appear as: "***Cannot read information off inserted buffer bottle***"

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-e7bd7023201bd310caa4cc18215a7aae2a3b4b82%2Fimage2.jpg?alt=media\&token=c5529c0c-7d94-44e6-9e2c-cd7c7b15bca9)

This second error may occur with the buffer cartridge, as sometimes during loading, the NextSeq 500/550 may read the wash buffer cartridge present in the instrument after loading the flow cell.

**Troubleshooting steps:**

1. Re-seat/reinsert consumable.
2. [Power Cycle](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001972.md) the instrument.
3. If issue persists, try a new consumable.
   * If resolved, it is most likely an RFID issue with the consumable. Contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) with the serial/lot number of the reagent to request a replacement.
4. If steps 1-3 do not resolve the issue, this indicates an issue with the RFID sensor on the instrument. Contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for further assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1914), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001914%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001914.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.