> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000002723.md).

# Single tile drop outs affecting metrics on the NextSeq 1000/2000

In some NextSeq 1000/2000 runs, it is possible to see a single tile or a small number of tiles in which the quality drops. If this occurs within the first 25 cycles, the passing filter can be negatively impacted within the tile(s). If this occurs at a later point in the run, it is possible to see small shifts in FWHM (the average full width of clusters at half maximum (in pixels)) or intensity.

**Cause:** The most common cause for this is a small bubble or bubbles present in the flow cell. Bubbles are more likely to affect the bottom surface but can sometimes impact both the top and bottom surface. It is most likely to see this with a small number of consecutive cycles at the beginning of the run.

Bubbles may result from improper reagent handling. Make sure each reagent cartridge is prepared for sequencing following recommended guidelines, see NextSeq 1000/2000 Product Documentation [Thaw the Bagged Cartridge](https://support-docs.illumina.com/IN/NextSeq10002000/Content/IN/NextSeq2000_1000/ThawBaggedCartridge_fNS.htm?Highlight=reagent%20preparation). For any runs impacted with single tile dropout, review the run as a whole for meeting output specifications ([Specifications for the NextSeq 1000/2000](https://www.illumina.com/systems/sequencing-platforms/nextseq-1000-2000/specifications.html)). Single tile dropouts are *very* unlikely to cause a run to fail to meet specifications. If run does not meet yield or Q30 specifications, evaluate for other potential causes.

**Identification:**\
In this example, the drop out occurred during the first 25 cycles which caused the tile to not pass filter.\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-ad9a0de35112fbb4bd01d110071bd87602193e41%2Fimage1.jpg?alt=media\&token=3ff37b35-3c50-4ebb-aa9e-dab7add1f087)

(Dark blue tiles can indicate much lower or zero values compared to other tiles.)

Effects of the bubble can be observed as an intensity drop out for a few cycles, followed by intensity recovering.\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-d7a63ab21fe0c53de070051814e508e96fe60900%2Fimage2.jpg?alt=media\&token=10a49a56-ada9-4497-b179-0175662909b6)

This is an additional example of a single tile drop-out occurring later in the run. This example presented as small FWHM spikes.\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-dc7f9e5ddc55005e1ff46b9f54c93bf3edd8c276%2Fimage3.jpg?alt=media)

FWHM for the given tile at cycle 72 higher than surrounding tiles, recovers in the subsequent cycle.\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-7500d4b614847d1a1b2058d3a5a2feeb90118e60%2Fimage4.jpg?alt=media)

Q30 was not calculated for the impacted tile.\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-4605879c175c5343a5e7f469a8f722e12d9afe35%2Fimage5.jpg?alt=media)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2723), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002723%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000002723.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.