# Pre Run check failed with an alignment check failure on the NextSeq 1000/2000

Error notification in Control Software: "An alignment check failed. Reload the flow cell and retry the pre-run check. If the problem persists, contact Illumina Technical Support."

![](/files/YshU920eNAukTEfGMUBk)

If this error occurs during the dry pre-run checks, the consumables are **not** marked as used.

**Troubleshooting steps**:

* If this error occurs, reseat the flow cell and repeat the pre-run checks.
* If error persists, power cycle the instrument [How to power cycle the NextSeq 1000/2000](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000003038)
* If error persists, attempt pre-run checks with a different flow cell.
  * If resolved by a new flow cell, the flow cell quality is the suspected root cause. Provide the lot and serial number to [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for replacement.
* If error persists with alternate flow cell, collect the following Logs and contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for file transfer link and next steps:
  * NextSeq 1000/2000 control software logs: /usr/local/Illumina/logs/nextseq1k2k-control-software
  * Pre-run check logs: usr/local/Illumina/Runs/Setup/\[Date\_Time]

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #3529), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000003529%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000003529.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.