Error notification in Control Software: "An alignment check failed. Reload the flow cell and retry the pre-run check. If the problem persists, contact Illumina Technical Support."
If this error occurs during the dry pre-run checks, the consumables are not marked as used.
Troubleshooting steps:
If this error occurs, reseat the flow cell and repeat the pre-run checks.
If error persists, attempt pre-run checks with a different flow cell.
If resolved by a new flow cell, the flow cell quality is the suspected root cause. Provide the lot and serial number to Illumina Technical Support for replacement.
If error persists with alternative flow cell, collect the following Logs and contact Illumina Technical Support for file transfer link and next steps:
NextSeq 1000/2000 control software logs: /usr/local/Illumina/logs/nextseq1k2k-control-software
Pre-run check logs: usr/local/Illumina/Runs/Setup/[Date_Time]