> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000009048.md).

# Troubleshooting System File errors during pre-run checks on the NextSeq 1000/2000

**Background**

During Pre-Run Checks, the NextSeq 1000/2000 verifies the Recipe files on instrument are compatible with the version of the reagent cartridge and flow cell that have been used to start the sequencing run.

If there are issues with locating or compiling the Recipe files, the following error will present during the Pre-Run Checks:

"Pre-Run Checks failed due to a system file issue. Contact Illumina Technical Support."

**Resolution steps**

If this error presents during the pre-run checks, this is often due to the Control Software not being compatible with the particular kit type. To resolve the issue, update the NextSeq 1000/2000 Control Software Suite to the latest version. See the following page for control software downloads.

[NextSeq 1000/2000 Control Software Suite Downloads](https://support.illumina.com/sequencing/sequencing_instruments/nextseq-1000-2000/downloads.html)

See the following Knowledge article for guidance on storing the reagent cartridge and flow cell during the upgrade process.

[Thawing and reagent preparation guidelines for the NextSeq 1000/2000 reagents](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000002430.md)

Once the Control Software upgrade is complete, proceed with setting up a new sequencing run using the same consumables (as long as they are within the recommended storage time after thawing). See [Returning NextSeq 1000/2000 reagents to storage](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000003033.md) for more information on storage time for opened consumables.

If the Control Software Suite has already been upgraded or the issue persists after upgrade, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html) for further assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9048), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009048%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000009048.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.