# Troubleshooting Failed to execute connectivity check input parameters on the NextSeq 1000/2000

The â€œFailed to execute connectivity check, input parameters are invalid" error can appear when there are network or internet connection issues with the selected output folder. This error occurs when selecting the output folder from **Settings** menu of the NextSeq 1000/2000 control software.

![](/files/k0YCkkPNesGejRoCsrBx)

**Potential Causes include:**

* Invalid Output folder.
* BaseSpace Hosting Location is not selected.

**Troubleshooting steps:**

1. Verify the output folder selected is a valid folder.

* See Knowledge Article: **Networking and output locations on the NextSeq 1000/2000**.

2. Select a Hosting Location from the **Connectivity** > **Hosting Location** Dropdown menu.

If unable to resolve the error, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for further guidance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #5227), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000005227%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000005227.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
