> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000008688.md). Agents can also search and read this documentation through its [MCP server](https://knowledge.illumina.com/~gitbook/mcp).

# Troubleshooting DRAGEN version warnings and errors after the NextSeq 1000/2000 Control Software upgrade to version 1.7

Then NextSeq 1000/2000 control software version 1.7 is only compatible with DRAGEN versions 4.2 and later. As a result, after the control software upgrade, previous versions of DRAGEN must be deleted in order to install the most up-to-date version. When the control software is initialized, the following warning message will be displayed as a reminder of the required workflow upgrades:

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-d246f625e4f229358a516acd69cbc0aa68f0af17%2Fimage1.jpg?alt=media)

To upgrade the DRAGEN workflows, all current DRAGEN workflows must be deleted. This can be done by:

1. Select the menu in the top left corner and selecting DRAGEN
2. Under version, select the trash can and verify the deletion of each workflow. (DRAGEN BCL Convert will need to be deleted last)
3. When all workflows are deleted, install the latest version of DRAGEN by selecting "Check Online" or "Browse for New Version" and following the prompts. (DRAGEN BCL Convert must be installed first)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-30848b59c598155b54439ebc3b3cd9755d88de93%2Fimage2.jpg?alt=media)

If newer versions of DRAGEN are installed prior to the deletion the older version of DRAGEN BCL Convert, an error will occur stating "Unable to delete the workflow for current DRAGEN version" when trying to delete the older version of DRAGEN. To resolve this, the user will need to delete all DRAGEN workflows and install them again starting with DRAGEN BCL Convert.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #8688), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000008688%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000008688.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.