> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-troubleshooting-list/000002545.md).

# Troubleshooting NovaSeq 6000 Control Software fails to initialize with Universal Copy Service error

**Background and Potential Root Causes**

If the NovaSeq 6000 Control Software is unable to establish communication with Universal Copy Service (UCS), the control software will fail to initialize.This can occur if there are runs that have stalled in the Cleaning Up state or if the UCS status was left pending prior to a control software upgrade.

In certain control software versions, this may occur at 97% initialization and typically presents the following error messages:

* **Universal Copy Service is unable to copy run files to the output folder location. Check that all network cables are connected or contact Technical Support for help**.
* **Unable to communicate with Universal Copy Service. Exit application, then use Windows Task manager as administrator to close UCS before restarting.**

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-ca6f862e28734668da66c68cd03c3578e8b65612%2Fimage1.jpg?alt=media\&token=78aec367-0276-43d1-a67c-58a5604df9c8)

**Note**: Error "**Universal Copy Service is unable to copy run files to the output folder location. Check that all network cables are connected or contact Technical Support for help**." can also occur *mid run* if connection to the output location is lost. If this is the case, please check the connection to network output location. File transfer will automatically restart once network connection is restored.

If initialization issues related to Universal Copy Service are observed, follow the Troubleshooting Actions below to either archive or delete the UCS Runs folder:

**Troubleshooting Actions**

1. Close the NovaSeq 6000 Control Software.
   1. Select the dropdown menu in the top-left corner, then select **Exit Application**.
2. Open Windows Task Manager
3. Navigate to the **Process** tab, and review process list for `ucs.exe` or `universal-copy-service.exe`. If Universal Copy Service (UCS) is running, right-click the process and select **End Process**.
4. Open `C:\ProgramData*\Illumina\Universal Copy Service\Runs`,
   * Option 1: Archive the existing `Runs` folder by moving the contents of this folder into a new "Archive" folder.
   * Option 2: Delete the contents of the `Runs` folder.
5. Power cycle the NovaSeq 6000. See the Knowledge Base article [How to Power Cycle the NovaSeq 6000 System](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000002470.md) for additional details.
6. After instrument has rebooted, restart the NovaSeq 6000 Control Software.

\***Note**: The `ProgramData` folder may be hidden. To un-hide this folder, select the **View** tab in File Explorer and then select the checkbox next to **“Hidden items”**.

If the run data needs to be manually transferred to the network output folder, see the following instructions:

1. On the NovaSeq 6000 control computer, open File Explorer and navigate to the run folder under `Z:\outputfolder\[Run_Folder]`
2. Copy the entire contents of the run folder from the `Z:\` drive to the intended network output location.
3. Within File Explorer, navigate to the run folder under `C:\ProgramData\Illumina\NovaSeq\NovaSeqTemp\[Run_Folder]\`
   1. **Note:** The `ProgramData` folder may be hidden. To un-hide the folder, select the **View** tab in File Explorer and then select the checkbox next to **“Hidden items”**.
4. Within the `[Run_Folder]` under the `ProgramData` directory, copy the following 2 files into the intended network output location:
   1. `RunParameters.xml`
   2. `SampleSheet.csv`

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2545), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002545%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-troubleshooting-list/000002545.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.