Troubleshooting a Failed Camera Board on the NextSeq 500/550

Background

The following error can occur at the beginning of a sequencing run or while the run is progressing, and causes the run to stop:

"All cameras disabled Originally caused by: Failed Camera Board."

Possible causes:

  • Transient communication issues.

  • A loose connection from the camera board.

  • A hardware failure at Main Printed circuit Assembly (MPCA).

Troubleshooting steps:

  1. Power cycle the NextSeq 500/550 following the instructions in How to Power Cycle the NextSeq 500/550

  2. If the system initializes without errors, run the Flow Cell Registration Test.

  3. If the Flow Cell registration Test passes without errors, proceed with setting up the next run and monitor the run performance.

  4. If there are initialization issues, the Flow Cell registration Test fails or a similar error appears at any stage, contact Illumina Technical Support.

For any feedback or questions regarding this article (Illumina Knowledge Article #1683), contact Illumina Technical Support techsupport@illumina.com.