> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-troubleshooting-list/000001909.md).

# Troubleshooting FPGA RFID ERROR STATUS 2 errors on the MiSeq

The FPGA mediates communication between subsystems of the MiSeq.

* If communication is lost with a subsystem, MiSeq Control Software (MCS) will show an error message.
* For the FPGA RFID ERROR STATUS 2 message, the error may occur if the PR2 bottle is removed from the instrument before "start wash" is selected.
* This error can indicate communication was lost or an issue with the instrument hardware (radio-frequency identification (RFID) reader).

**Error:** The FPGA reported an error while executing command "*RFID3BLKWR 1 19 80000002*"; error message: "*RFID ERROR STATUS 2 A00000*" (*Illumina.Instrument.Bolt.Logic.Mainboard.FPGAErrorResponseException*)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-c9f5bcab3ab5d51fffe53f47c20affd03efa276c%2Fimage1.png?alt=media\&token=73678cd0-5354-4a5d-be65-db6ec8e8c5ea)

**To troubleshoot:**

1. Verify run completed successfully.
2. Complete a post-run wash if it has not been completed.
3. Power cycle the MiSeq following the guidance in the Knowledge Base articles [How to Power Cycle the MiSeq](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001894.md) and[How To Power Cycle the MiSeq Video](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000005871.md).
4. If the instrument initializes without error, continue using the MiSeq for sequencing.
5. If the error persists, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/) for assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1909), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001909%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-troubleshooting-list/000001909.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.