Viewing NextSeq 1000/2000 runs in Sequencing Analysis Viewer
Background
The latest available version of Sequencing Analysis Viewer (SAV) is v3.0. SAV v3.0 has Windows, Mac, and Linux versions; however, the current Linux version is only compatible with Enterprise Linux versions (EL8 and above). The NextSeq 1000/2000 currently has two OS versions. The older CentOS 7 (with NCS v1.7.x and older) that is EL7 and is not compatible. The newer OS version is Oracle Linux 9 (with NCS 2.x) that is EL9 is compatible.
For more information, see Knowledge article Common Questions Related to Sequencing Analysis Viewer (SAV) software and the NextSeq 1000/2000. Note: for more details on NextSeq1000/2000 data with ealier SAV versions, see section below.
The %Occupied metric has not been optimized for the reduced nanowell pitch width on the NextSeq 1000/2000. %Occupied is not a reliable indicator of flow cell loading on the NextSeq 1000/2000.
%Loading Concentration can be used as a gauge for how loaded the flow cell is. For more information on %Loading Concentration see Knowledge Base Optimizing loading concentration on the NextSeq 1000/2000.
**Schematic of Flow cells with tiles
P1 Flow cell**



P2 Flow cell



P3 Flow cell (similar for P4 Flow cell)


P3 Flow cell have 14 tiles per swaths and 6 swaths per tile.
P4 Flow cell have 16 tiles per swaths and 6 swaths per tile.
Tile Naming Convention
The tile name is a four-digit number that represents the tile position on the flow cell. For example, tile name 1205 indicates top surface, swath 2, tile 05.
The first digit represents the surface: 1 for top or 2 for bottom.
The second digit represents the swath number: 1, 2, 3, 4, 5, or 6.
The last two digits represent the tile number. For swath numbers 1-4, the numbering starts with 01 at the outlet end of the flow cell through 11 at the inlet end. For swath numbers 5-6, the numbering starts with 01 at the inlet end and 11 at the outlet end.
Sequencing Analysis Viewer v2.5.12 and below
Version 2.5.12 of Sequencing Analysis Viewer is required for reviewing runs generated on NextSeq 1000/2000 control software version 1.5 or later.
SAV v2.5.12 is only validated for use with NextSeq 1000/2000 and is not required for any other platform.
The flow cells for the P1 and P2 600 cycle kits have not changed, but the way the RunInfo and ErrorMetrics are loaded in SAV has been updated
Example of Error displayed when loading run from NextSeq 1000/2000 control software v1.5 into SAV version 2.4.7 or earlier.