Troubleshooting Local Run Manager originally caused by an internal server error on NextSeq 500/550

The "Local Run Manager: Originally caused by an internal server error" notification in the NextSeq 500/550 Control Software (NCS) can indicate either a transient communication issue or that the PostgreSQL service is not running.

Troubleshooting steps:

  1. Power cycle the NextSeq 500/550 following the instructions in How to Power Cycle the NextSeq 500/550.

  2. If the NCS reinitializes without errors, continue with next sequencing run.

  3. If the notification persists in the NCS after a power cycle, restart the PostgreSQL service using the following steps:

    1. Close the NCS.

    2. Search for the Services App in Windows task bar, then right-click and select Run as Administrator. Enter the sbsAdmin or other local windows admin-level account password

    3. Search for PostgreSQL and check to see if it is 'Running'. If not, right-click and select Start.

    4. Relaunch the NCS.

If the issue persists, contact Illumina Technical Support for additional assistance.

For any feedback or questions regarding this article (Illumina Knowledge Article #7806), contact Illumina Technical Support techsupport@illumina.com.