Troubleshooting Local Run Manager originally caused by an internal server error on NextSeq 500/550
The "Local Run Manager: Originally caused by an internal server error" notification in the NextSeq 500/550 Control Software (NCS) can indicate either a transient communication issue or that the PostgreSQL service is not running.
Troubleshooting steps:
Power cycle the NextSeq 500/550 following the instructions in How to Power Cycle the NextSeq 500/550.
If the NCS reinitializes without errors, continue with next sequencing run.
If the notification persists in the NCS after a power cycle, restart the PostgreSQL service using the following steps:
Close the NCS.
Search for the Services App in Windows task bar, then right-click and select Run as Administrator. Enter the sbsAdmin or other local windows admin-level account password
Search for PostgreSQL and check to see if it is 'Running'. If not, right-click and select Start.
Relaunch the NCS.
If the issue persists, contact Illumina Technical Support for additional assistance.
For any feedback or questions regarding this article (Illumina Knowledge Article #7806), contact Illumina Technical Support techsupport@illumina.com.