# Troubleshooting an Async Firmware Operation Error on the NextSeq 500/550

This error is due to a transient communication issue that does not affect run progression or performance.

**Possible error messages:**\
"AsyncFirmwareOperationCommand timed out waiting for started after XX:XX:XX."\
"While waiting on an asynchronous firmware operation: camera bitmask 111111 Originally caused by: Failed XY Stage."\
![](/files/NdxE7JOiU5XPKLJjGVxI)

**Troubleshooting steps:**

1. If the run is continuing, allow the run to complete and monitor.
2. After the run completes, review run performance. See [Specifications for the NextSeq 500/550 System](https://www.illumina.com/systems/sequencing-platforms/nextseq/specifications.html).
3. Power cycle NextSeq 500/550.\\\*
4. If the system initializes, proceed with the next sequencing run.

If the same notification is displayed over 2-3 consecutive runs, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support), for further assistance.

\\\*For further information, search for Knowledge Base Article **How to Perform a NextSeq 550 Power Cycle.**

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1680), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001680%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001680.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.