> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-reference_material-list/000001882.md).

# Air Filter replacement on the iSeq 100

The Air Filter on the iSeq 100 is a single-use foam piece that covers the two fans on the back of the instrument; it ensures proper cooling and prevents debris from entering the system. The Air Filter must be replaced every 6 months. The instrument generates a pop-up notification indicating it is time to change the Air Filter. The instrument ships with one installed and one spare. Keep the Air Filter in its package until use.

If the instrument is under an active Illumina service contract, Illumina Technical Support (TS) can provide an iSeq 100 Spare Air Filter as needed. When the instrument is not under warranty, replacements are user-supplied.

iSeq 100 Spare Air Filter catalog number for ordering: **20023928.**

**How to Replace an iSeq Air Filter:**

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-a5d05f3f54da2baf03fdac677bab0c2582833821%2Fimage1.png?alt=media\&token=cc1ec7ce-66d5-43f8-8d6c-12553180340b)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-9ca2b63f7ef62cff5faa5049f68c0c4e9c681f9c%2Fimage2.png?alt=media\&token=56741e6a-32fa-4f2f-a17f-99a14146dee5)

More information, reference [iSeq 100 Sequencing System Guide](https://support-docs.illumina.com/IN/iSeq100/Content/IN/iSeq/Maintenance.htm)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1882), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001882%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-reference_material-list/000001882.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.