> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miniseq/instrumentation-miniseq-troubleshooting-list/000008099.md).

# Troubleshooting RFID block can not be written (due to lock status) Error on the MiniSeq

Users may encounter the following error at the end of a sequencing run on the MiniSeq:

*"Exception generated from IxCmd Illumina.Spark.EmbeddedCommandSet.ix\_rfid\_tag\_set : RFID block can not be written (due to lock status)"*

This error message may be generated during the final cycle of a run. It does not represent an issue with sequencing performance or quality, and is typically indicative of a communication error onboard the instrument.

**Troubleshooting steps:**\
To proceed, acknowledge the error and power cycle the instrument as follows:

1. From home screen, select **Manage Instrument**.
2. Select **Shutdown Options**.
3. Select **Shut Down**.
4. Wait 1-2 minutes before turning on instrument with button on front left of instrument.
5. Wait for the Windows Operating System and MiniSeq Control Software to fully initialize.

Users may confirm that their run completed by navigating to their output folder and confirming the presence of the "CopyComplete.txt" flag file. If this file is not present within 10 minutes after power cycling, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for further assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #8099), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000008099%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miniseq/instrumentation-miniseq-troubleshooting-list/000008099.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.