> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000002826.md).

# Maximum read length for Illumina sequencing platforms

**Maximum supported read length**

Illumina platforms and sequencing kits allow for a variety of [sequencing applications](https://www.illumina.com/science/technology/next-generation-sequencing/plan-experiments/read-length.html). To ensure the highest level of quality, Illumina supports reads up to a certain length depending on the sequencing platform and SBS kit version; refer to [How many cycles of SBS chemistry are in my kit?](/instrumentation/general/instrumentation-general-reference_material-list/000007002.md).

It is possible to choose a longer read length during run setup in [Local Run Manager (LRM)](https://emea.support.illumina.com/bulletins/2018/04/local-run-manager-2--how-to-use-a-custom-library-prep-kit.html), [BaseSpace Run Planning](https://help.basespace.illumina.com/sequence/plan-runs), or the instrument Control Software. However, when the read length exceeds the supported length, Illumina cannot guarantee maintaining high data quality.

**Table 1.** Maximum supported read length for sequencing platforms and SBS reagent kits.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-eddcbc301eca8c0cefb99f40e9b7417e85c57740%2Fimage1.png?alt=media)

\* MO: Mid-output / HO: High-output

**Maximum read length for index reads**

* Instrument control software: enter custom Index Read length during run setup for certain instruments, as shown below in Table 2.
* LRM v2: enter up to 100 characters for the maximum index cycles (see [How to use a custom library prep kit in Local Run Manager v2](/software/local-run-manager/software-local-run-manager-reference_material-list/000001317.md) for additional information on custom libraries with Local Run Manager).
* LRM v3: enter up to 20 characters, or use manual mode to allow for between 20 and 100 bp.
* LRM v4: enter up to 20 characters, or use manual mode to allow for between 20 and 1000 bp.
* To ensure optimal sequencing quality for the index read, do not exceed the supported read length.

**Table 2.** Maximum individual index length for sequencing platforms.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-a276c0f23adab2202397018c262c9619f915c120%2Fimage2.png?alt=media)

\*When setting up runs in Cloud or Sample Sheet mode, index length is limited by the BCL Convert limitations listed below.

\*\*For index length between 20-1000 bp, use manual mode as LRM v4 is limited to 20 bp.

\*\*\*index length is not limited for runs set up in manual mode without analysis.

**Note**: BCL Convert has its own index limitations. BCL Convert 4.2 or lower is limited to 27 index cycles total. For BCL Convert 4.3 and higher, the maximum index length has been increased to 27 cycles for each index read.

If using the bclconvert override cycle settings, bases specified for UMIs (U) in the index reads do not count against the maximum index length.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2826), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002826%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000002826.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.