> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009205.md).

# Troubleshooting Memory Metal Alloy (MMA) Motion Stalls on the MiSeq i100 Series

**Background**

The Memory Metal Alloy (MMA) assembly is responsible for opening and closing valves when directing the flow of reagents through the dry cartridge. Issues with one of the MMA valves not responding will trigger the instrument to stop the run and present a Sequencing failed error:

"Failed executing Protocol 2Read2Index after \<Time Stamp>. Motion stall \<Date Stamp>"

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-bfecaa6295998e533dcfa9197c0ad7237de9ccb1%2Fimage1.png?alt=media)

**Troubleshooting Actions**

1. Select **Eject Cartridge**, then remove the dry and wet sequencing cartridges from the instrument.
2. Power cycle the instrument to ensure all board communications are reset.
   1. See Knowledge Article [How to Power Cycle the MiSeq i100 Series](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009277.md) for more information.
3. Run the instrument System Health Checks using the Test Cartridges provided with the instrument.
   1. See Knowledge Article [How to Setup the System Checks on the MiSeq i100 Series Instruments](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009206.md) for more information.
4. Once the System Health Checks are complete, select **Clear Checks** and then proceed to the Control Software home screen.
5. Follow the Export Logs - Standardworkflow to bundle the System Health Check results.
   1. Select the affected Sequencing Run from the list when setting up the Export Logs.
   2. See Knowledge Article [How to Export Logs on the MiSeq i100 and NovaSeq X Series Instruments](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000009265.md) for more information.
6. Once the Export Logs workflow has completed and transferred to the External Storage, contact [Illumina Tech Support](https://www.illumina.com/company/contact-us.html) for further steps.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9205), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009205%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009205.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.