> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-troubleshooting-list/000001306.md).

# Troubleshooting "Unable to Generate a Focus Model" error on the NovaSeq 6000

**Error message**:

"Unable to generate a focus model for the flow cell, which requires an onsite correction. Contact Technical Support for help."

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-06956c99fd7b1c7034e33bbd039e6efc7318ffa3%2Fimage1.jpg?alt=media\&token=d881adb0-cd18-4cd7-aadb-37c9f909687f)

**Figure 1.** Example Error message in NovaSeq Control Software

The focus model is generated by the Focus Tracking Module (FTM) and Line Generation Module (LGM). The FTM is a secondary onboard laser diode separate from main Red and Green lasers of the LGM. Failure to generate a focus model is related to either low intensities of the lasers on the Focus Tracking Module (FTM), a consumable issue with the flow cell, fluidics, temperature, or other optical issues.

**Troubleshooting steps:**

1. Power cycle the instrument.
   1. For more information, review the Knowledge Base article [How to Power Cycle the NovaSeq 6000](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000002470.md).
2. Once the instrument initializes successfully, perform a system check.
   1. For more information, review the Knowledge Base article [How to perform system checks on the NovaSeq 6000](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000002495.md).
3. Inspect the flow cell for any debris/glass shards or defects.
4. Collect the following files and contact [Illumina Technical Support](https://emea.illumina.com/company/contact-us.html#/united-states/technical-support) (<techsupport@illumina.com>) for instructions on how to upload the compressed Alignment folder to a secure site for review and next steps.
   1. Logs folder: C:\ProgramData\Illumina\NovaSeq\NovaSeqTemp\[Run\_Folder]\Logs
   2. Alignment folder: C:\ProgramData\Illumina\NovaSeq\NovaSeqTemp\[Run\_folder]\Alignment

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1306), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001306%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-troubleshooting-list/000001306.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.