> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000001530.md). Agents can also search and read this documentation through its [MCP server](https://knowledge.illumina.com/~gitbook/mcp).

# How to set the output location in the NovaSeq 6000 Control Software

**Background**

The NovaSeq 6000 platform requires the designation of an output folder path to a network location prior to the start of a sequencing run in Manual mode. For more information on how to configure the output folder path in File-Based mode, refer to the [NovaSeq 6000 Sequencing System Guide](https://support.illumina.com/sequencing/sequencing_instruments/novaseq-6000/documentation.html).

**Important Notes**

* If the run uses BaseSpace Sequence Hub for both analysis and storage, setting a network output folder location is optional. In this scenario a warning will be presented "The output folder must exist and cannot be on C:\ D:\ or Z:\ drives". After importing a valid Sample Sheet, the "Review" option will be available to move forward with setting up the run.
* A mapped network drive should be configured by your network administrator.
* The UNC path for the network drive is required for setting the output location. Using a mapped letter, such as Y:, results in an Invalid Drive error message.
  * Example of a UNC path: \servername\share-directory\folder1\\

**To specify a default output location in the NovaSeq Control Software:**

1. From the **Main Menu**, select **Settings**.
2. Select **Manual**.
3. Select **Browse** (next to the Output Folder field).
4. Type or copy and paste the UNC path into the address bar.
5. Select **Enter** on keyboard, and input network credentials if prompted.
6. Designate the desired folder and click **Select Folder**.
7. Select **Save**.

Once the default output location has been designated, the output folder field will automatically populate on the Run Setup screen for future runs. To change the output location for a single run, follow steps 3-7 on the Run Setup screen.

For additional information on networking requirements or run setup procedures, refer to the following resources:

[NovaSeq 6000 Sequencing System Guide](https://support.illumina.com/sequencing/sequencing_instruments/novaseq-6000/documentation.html)\
[NovaSeq 6000 Sequencing System Site Prep Guide](https://support.illumina.com/downloads/novaseq-site-prep-guide-1000000019360.html)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1530), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001530%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000001530.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.