Wash Error "Binary Part number revision .. byte1 is invalid" on the MiniSeq
When performing a wash using a wash flow cell or wash tray, users may see the following error:
Error language: Binary Part number revision (0xBEFF) byte 1 (0xBE) in invalid, must be either 0x00, or ASCII for A-Z or 0-9 Parameter name: byte1

This error occurs when attempting to set up a wash using either a wash tray or used flow cell that was previously used to perform a Radio-frequency identification (RFID) check in the MiniSeq Service Software (MnSS). This is typically encountered following an Illumina Field Service Engineer (FSE) visit, when the consumables or wash tray were used during MnSS RFID check test during troubleshooting. This is resolved by changing the RFID status for "Block Read 1" in the MnSS. The RFID status for "Block Read 1" should contain all "1". If any values are set to "0", they must be switched back using the RFID check-in MnSS.
Troubleshooting steps:
Exit the MiniSeq Control Software (MnCS): select Manage Instrument → Shutdown Options → Exit to Windows.
Open the MiniSeq Service Software (MnSS).
Select Mechanical & Fluidics Tests.
Select RFID.
Select Initialize, an icon of two arrowheads chasing each other at top of MnSS. Allow the consumables and to system initialize.

The above image shows both correct and incorrect flow cell and reagent cartridge results for this test. Verify which reagent is incorrectly labeled.
Re-run the test with the component that has the wrong RFID block (Block Read 1 all "0"). This will "flip" it back to the correct settings.
Select Home → Exit to MiniSeq Control Software
Once MnCS initializes, set up a wash with the consumables. If just the flow cell was switched, MnCS will recognize the flow cell as a used flow cell and allow the wash to proceed.
For any feedback or questions regarding this article (Illumina Knowledge Article #1222), contact Illumina Technical Support techsupport@illumina.com.
Last updated
Was this helpful?
