> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009380.md).

# Troubleshooting 'Status code 523' Errors in Cloud Settings on the MiSeq i100 Series

**Background**

The MiSeq i100 Series Control Software leverages HTML-based software design to serve up user interface (UI) screens. Disruptions to the instrument network connection can prevent normal UI loading and present the following error messages in a banner at the top of the UI:

`'An unexpected error occurred. Details: Status code 523'`

**Troubleshooting Actions**

1. Confirm that the ethernet cable is plugged into the instrument and is connected to the LAN1 port at the back of the instrument.
2. Review the **Network Settings** page for potential configuration issues:
   1. Log into the Control Software as an **Administrator** user.
   2. From the Home screen, select the dropdown menu in the top-left corner, then select **Settings**.
   3. Select **Network Settings**.
   4. Confirm that LAN1 shows the expected IP address, gateway, and DNS IP address information for the local network connection.
      1. If any values are missing or incorrect, consult with local IT team and update values accordingly.
   5. Select **Save**.
3. Navigate to the **Cloud Settings** menu to confirm that error message is resolved.

If the banner presenting the '`Status code 523`' persists after confirming Network Settings are correct, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9380), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009380%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009380.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.