Procedures for cleaning flow cells for Illumina sequencers
iSeq 100
No cleaning required.
Only touch the plastic when handling the flow cell.
Avoid touching the electrical interface, CMOS sensor, glass, and gaskets on either side of the glass.
Reference: iSeq 100 Sequencing System Product Documentation.
MiniSeq
Put on a new pair of powder-free gloves.
Remove the flow cell from the container.
Clean the glass surface of the flow cell with a lint-free alcohol wipe.
Dry with a lint-free lens cleaning tissue. Use care around the black flow cell gasket.
Inspect the flow cell port for obstructions. Make sure that the gasket is well-seated.
Reference: MiniSeq System Guide.
MiSeq
Put on a new pair of powder-free gloves.
Remove the flow cell from the container.
Lightly rinse the flow cell with laboratory grade water until both the glass and the plastic cartridge are thoroughly rinsed of excess salts.
Using care around the black flow cell port gasket, thoroughly dry the flow cell glass and plastic with a lint-free lens cleaning tissue. Gently pat dry in the area of the gasket and adjacent glass.
Clean the flow cell glass with an alcohol wipe. Make sure that the glass is free of streaks, fingerprints, and lint or tissue fibres.
DO NOT USE THE ALCOHOL WIPE ON THE FLOW CELL PORT GASKET.
Dry excess alcohol with a lint-free lens cleaning tissue.
Inspect the flow cell port for obstructions. Make sure that the gasket is well-seated. If the gasket appears to be dislodged, gently press it back until it sits securely around the flow cell ports.
Reference: MiSeq System Guide for Windows 10 and MiSeq System Guide for Windows 7.
NextSeq 500/550
Remove the flow cell from the foil packaging.
Open the clear plastic clamshell package and remove the flow cell.
Clean the glass surface of the flow cell with a lint-free alcohol wipe. Dry the glass with a low-lint lab tissue.
NextSeq 1000/2000
The flow cell MUST NOT be cleaned.
Flow cells can look streaky; they are fine to use.
Reference NextSeq 1000/2000 Product Documentation.
HiSeq 2500
Remove the clustered flow cell from the container using a pair of plastistats.
Rinse the flow cell with laboratory-grade water. Dry with a lens cleaning tissue.
Fold an alcohol wipe to approximately the size of the flow cell.
Hold the edges of the flow cell with 2 fingers so that the inlet and outlet ports face up.
Wipe each side of the flow cell using a single sweeping motion. Repeat until clean, refolding the alcohol wipe with each pass.
Dry with a dry lens cleaning tissue.
Protect the flow cell from dust until ready to load it onto the instrument.
Reference: HiSeq 2500 System Guide.
HiSeq 3000/4000/X
It is not necessary to clean the flow cell.
If desired, can clean the glass surface of the flow cell with a lint-free alcohol wipe. Dry the glass with a low-lint lab tissue.
NovaSeq 6000
It is not necessary to clean the flow cell.
If desired, clean the glass surface of the flow cell with a lint-free alcohol wipe. Dry the glass with a low-lint lab tissue.
Reference: NovaSeq 6000 Sequencing System Guide.