> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-faq-list/000005147.md).

# Procedures for cleaning flow cells for Illumina sequencers

**iSeq 100**

* No cleaning required.
* Only touch the plastic when handling the flow cell.
* Avoid touching the electrical interface, CMOS sensor, glass, and gaskets on either side of the glass.
* Reference: [iSeq 100 Sequencing System Product Documentation](https://support.illumina.com/downloads/iseq-100-system-guide.html).

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-1470258480a2cb1c0f43395ba68c472d3e3b4755%2Fimage1.png?alt=media)

**MiniSeq**

* Put on a new pair of powder-free gloves.
* Remove the flow cell from the container.
* Clean the glass surface of the flow cell with a lint-free alcohol wipe.
* Dry with a lint-free lens cleaning tissue. Use care around the black flow cell gasket.
* Inspect the flow cell port for obstructions. Make sure that the gasket is well-seated.
* Reference: [MiniSeq System Guide](https://support.illumina.com/downloads/miniseq-system-guide-1000000002695.html).

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-0ebe9f15df7deb444f1d0dd706dd0938efafb354%2Fimage2.png?alt=media)

**MiSeq**

* Put on a new pair of powder-free gloves.
* Remove the flow cell from the container.
* Lightly rinse the flow cell with laboratory grade water until both the glass and the plastic cartridge are thoroughly rinsed of excess salts.
* Using care around the black flow cell port gasket, thoroughly dry the flow cell glass and plastic with a lint-free lens cleaning tissue. Gently pat dry in the area of the gasket and adjacent glass.
* Clean the flow cell glass with an alcohol wipe. Make sure that the glass is free of streaks, fingerprints, and lint or tissue fibres.
* **DO NOT USE THE ALCOHOL WIPE ON THE FLOW CELL PORT GASKET.**
* Dry excess alcohol with a lint-free lens cleaning tissue.
* Inspect the flow cell port for obstructions. Make sure that the gasket is well-seated. If the gasket appears to be dislodged, gently press it back until it sits securely around the flow cell ports.
* Reference: [MiSeq System Guide for Windows 10](https://support.illumina.com/downloads/miseq_system_user_guide_15027617.html) and [MiSeq System Guide for Windows 7](https://support.illumina.com/downloads/miseq_system_user_guide_win7.html).

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-b47e7c8d7862d7ea9f0c6f9384f3fb10f4b18386%2Fimage3.png?alt=media)

**NextSeq 500/550**

* Remove the flow cell from the foil packaging.
* Open the clear plastic clamshell package and remove the flow cell.
* Clean the glass surface of the flow cell with a lint-free alcohol wipe. Dry the glass with a low-lint lab tissue.
* Reference: [NextSeq 500 and NextSeq 550 Sequencing System Guide](https://support.illumina.com/downloads/nextseq-500-550-systems-product-documentation.html)

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-17b481c239260642b9673d3f0b2d6c578e68d52a%2Fimage4.png?alt=media)

**NextSeq 1000/2000**

* The flow cell **MUST NOT** be cleaned.
* Flow cells can look streaky; they are fine to use.
* Reference [NextSeq 1000/2000 Product Documentation](https://support.illumina.com/downloads/nextseq-1000-2000-product-documentation-200027171.html).

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-473dc728b791f48e042304de9619406e633d4c26%2Fimage5.png?alt=media)

**NovaSeq 6000**

* It is not necessary to clean the flow cell.
* If desired, clean the glass surface of the flow cell with a lint-free alcohol wipe. Dry the glass with a low-lint lab tissue.
* Reference: [NovaSeq 6000 Sequencing System Guide](https://support.illumina.com/downloads/novaseq-6000-system-guide-1000000019358.html).

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-7f6443eaa575ce42393d62420d0467b5f62b2bea%2Fimage6.png?alt=media)

**NovaSeq X Series**

* Put on a new pair of powder-free gloves to avoid contaminating the glass surface of the flow cell.
* With the flow cell foil package over a flat surface, peel open the foil from the corner tab
* Remove the flow cell from the package. Grasp the flow cell by the sides to avoid touching the glass or the underside gaskets.
* **Always** clean the flow cell, whether or not a contaminant is present, using the following steps:
  1. Wet a Contec Polynit Heatseal wipe with isopropyl alcohol (70%).
  2. Gently clean the applicable surface. Wipe in a lengthwise direction (top to bottom direction) only, avoiding gaskets and manifolds.
  3. Repeat steps **a** and **b** until surfaces are clear.
  4. Dry with a new polynit heatseal wipe or an unused side of the used wipe to avoid contamination.
* Do NOT use other lint-free wipes such as Kim Wipes or similar products. Only use Contec Polynit Heatseal wipes.
* Always use isopropyl alcohol over ethanol due to higher purity.
* Reference: [NovaSeq X Series](https://support-docs.illumina.com/IN/NovaSeqX/Content/IN/FrontPages/NovaSeqX.htm).

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-35ba7c7cc413eb517e21f229472a093c587b62a6%2Fimage7.png?alt=media)

**MiSeq i100**

* No cleaning required (CMOS patterned flow cell integrated into the Dry cartridge).
* Cartridge should be gripped from the side to avoid touching the flowcell circuit/CMOS sensor.
* Reference: [MiSeq i100 Plus System](https://support-docs.illumina.com/IN/MiSeqi100Series/Content/IN/FrontPages/MiSeqi100.htm).

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-e72f3fb77095ab6b249b4334128de95b0c34221e%2Fimage8.png?alt=media)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #5147), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000005147%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-faq-list/000005147.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.