Procedures for cleaning flow cells for Illumina sequencers
Last updated
Was this helpful?
Last updated
Was this helpful?
iSeq 100
No cleaning required.
Only touch the plastic when handling the flow cell.
Avoid touching the electrical interface, CMOS sensor, glass, and gaskets on either side of the glass.
Reference: .
MiniSeq
Put on a new pair of powder-free gloves.
Remove the flow cell from the container.
Clean the glass surface of the flow cell with a lint-free alcohol wipe.
Dry with a lint-free lens cleaning tissue. Use care around the black flow cell gasket.
Inspect the flow cell port for obstructions. Make sure that the gasket is well-seated.
Reference: .
MiSeq
Put on a new pair of powder-free gloves.
Remove the flow cell from the container.
Lightly rinse the flow cell with laboratory grade water until both the glass and the plastic cartridge are thoroughly rinsed of excess salts.
Using care around the black flow cell port gasket, thoroughly dry the flow cell glass and plastic with a lint-free lens cleaning tissue. Gently pat dry in the area of the gasket and adjacent glass.
Clean the flow cell glass with an alcohol wipe. Make sure that the glass is free of streaks, fingerprints, and lint or tissue fibres.
DO NOT USE THE ALCOHOL WIPE ON THE FLOW CELL PORT GASKET.
Dry excess alcohol with a lint-free lens cleaning tissue.
Inspect the flow cell port for obstructions. Make sure that the gasket is well-seated. If the gasket appears to be dislodged, gently press it back until it sits securely around the flow cell ports.
Reference: and .
NextSeq 500/550
Remove the flow cell from the foil packaging.
Open the clear plastic clamshell package and remove the flow cell.
Clean the glass surface of the flow cell with a lint-free alcohol wipe. Dry the glass with a low-lint lab tissue.
NextSeq 1000/2000
The flow cell MUST NOT be cleaned.
Flow cells can look streaky; they are fine to use.
NovaSeq 6000
It is not necessary to clean the flow cell.
If desired, clean the glass surface of the flow cell with a lint-free alcohol wipe. Dry the glass with a low-lint lab tissue.
NovaSeq X Series
Put on a new pair of powder-free gloves to avoid contaminating the glass surface of the flow cell.
With the flow cell foil package over a flat surface, peel open the foil from the corner tab
Remove the flow cell from the package. Grasp the flow cell by the sides to avoid touching the glass or the underside gaskets.
Always clean the flow cell, whether or not a contaminant is present, using the following steps:
Wet a Contec Polynit Heatseal wipe with isopropyl alcohol (70%).
Gently clean the applicable surface. Wipe in a lengthwise direction (top to bottom direction) only, avoiding gaskets and manifolds.
Repeat steps a and b until surfaces are clear.
Dry with a new polynit heatseal wipe or an unused side of the used wipe to avoid contamination.
Do NOT use other lint-free wipes such as Kim Wipes or similar products. Only use Contec Polynit Heatseal wipes.
Always use isopropyl alcohol over ethanol due to higher purity.
MiSeq i100
No cleaning required (CMOS patterned flow cell integrated into the Dry cartridge).
Cartridge should be gripped from the side to avoid touching the flowcell circuit/CMOS sensor.
Reference:
Reference .
Reference: .
Reference: .
Reference: .