Troubleshooting a Software Shutdown Issue Warning on the NextSeq 500/550 with NCS 4.0.X

The following message can appear when shutting down or restarting the NextSeq 500/550. This can be related to having Proactive enabled.

Software shutdown issue(s): Data from the previous sequencing run is being sent to persistent storage. Select Cancel to wait for the transfer to finish. Do you want to proceed?

Troubleshooting steps:

  1. If the instrument not connected to the internet, disable Proactive monitoring as follows:

    1. Navigate to the NextSeq Control Software.

    2. Select System Settings.

    3. Deselect the option Send Instrument Performance Data to Illumina.

  2. If all runs have completed transfer and/or upload, apply instructions found in the following article: Troubleshooting a NextSeq 500/550 NextSeq Control Software (NCS) v4.0+ UCS error.

  3. If the most recent run is still being transferred and/or uploaded, it is acceptable to select Proceed and ignore the message.

  • If run data had not finished copying, Universal Copy Service (UCS) will continue process after instrument has been restarted.

  • Note: This warning will have no impact on run quality or performance.