Troubleshooting a Software Shutdown Issue Warning on the NextSeq 500/550 with NCS 4.0.X
The following message may appear when shutting down or restarting the NextSeq 500/550 and can be associated with Proactive being enabled. This warning does not affect run quality or system performance.
"Software shutdown issue(s): Data from the previous sequencing run is being sent to persistent storage. Select Cancel to wait for the transfer to finish. Do you want to proceed?"

Troubleshooting steps:
If the instrument not connected to the internet, disable Proactive monitoring as follows:
Navigate to the NextSeq Control Software.
Select System Settings.
Deselect the option Send Instrument Performance Data to Illumina.
If all runs have completed transfer and/or upload, apply instructions found in the following article: Troubleshooting Universal Copy Server error on the NextSeq 500/550 (NCS 4.0+).
If the most recent run is still being transferred and/or uploaded, it is acceptable to select Proceed and ignore the message.
If run data had not finished copying, Universal Copy Service (UCS) will continue process after instrument has been restarted.
For any feedback or questions regarding this article (Illumina Knowledge Article #2449), contact Illumina Technical Support techsupport@illumina.com.