> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miniseq/instrumentation-miniseq-reference_material-list/000006617.md).

# How to install custom recipes on MiniSeq Control Software v2.x

When using a custom recipe for run setup with the MiniSeq Control Software (MnCS) v2.x, the option to select the custom recipe will occur after the reagent cartridge is loaded (not during the run setup page) only. By this point, a new flow cell, buffer cartridge, and reagent cartridge are registered before selecting the custom recipe.

Contact the local Field Applications Scientist (FAS) for questions regarding a custom recipe or contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/).

To install a custom recipe on the MiniSeq, it must be copied to the custom recipes location: `C:\Program Files\Illumina\MiniSeq Control Software\Recipe\Custom\[High or Mid]`

After installing the custom recipe in the appropriate location, select "**Enable Custom Recipes"** in System Customization in MnCS. The Custom Recipe option will not appear during the run setup if not selected.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-3b18e6ef3eeb55e30de8fbc40c3ca57a4888236c%2Fimage1.png?alt=media\&token=9c7ab822-db56-4a45-ab51-8c683e5b38ec)

Select the custom recipe after completing the run setup and loading the consumables. Loading consumables will validate which recipes are available for that flow cell type.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-47f0b9654569341ee17993fe3b8a6879f30bf0d6%2Fimage2.png?alt=media\&token=78aef7a0-fbb1-4372-a376-49f829bc7534)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #6617), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000006617%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miniseq/instrumentation-miniseq-reference_material-list/000006617.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.