Offline DRAGEN License installation on the MiSeq i100 Series

Background

The DRAGEN Analysis software requires a valid license before use. For instruments that are connected to the internet, the license can be installed and validated from the DRAGEN License server.

For instruments that do not have internet connection, see the following steps.

Resolution steps

  1. Contact Illumina Customer Care for the Offline License file keyed to the specific MiSeq i100 Serial Number.

  2. Once available, transfer the zip/compressed license file to a USB 3.0 compatible storage device.

  3. Plug in the USB 3.0 storage device into the blue USB 3.0 port on the left side of the instrument.

  4. Login to the Control Software as an Administrator user account.

  5. Within the Control Software, use the External Storage user interface (UI) to add the USB storage device.

    • See the Add a USB Drive section of the MiSeq i100 Product Documentation for further details.

  6. From the Settings menu, select DRAGEN, then select the License tab.

  7. Select Update from file, then select From external storage.

  8. Navigate to the zip/compressed license file on the USB 3.0 storage device.

  9. Click Select.

  10. Allow the Control Software to install the license file.

  11. Select the Versions tab.

  12. For version line, select the ellipses (...) on the right side, then select Run self test.

  13. Once the Self Test completes, the instrument DRAGEN Analysis software is ready to use.

For any feedback or questions regarding this article (Illumina Knowledge Article #9557), contact Illumina Technical Support techsupport@illumina.com.