Specifications for the NovaSeq 6000
**Performance Parameters for NovaSeq 6000 Reagent Kits
Total Expected Output for Reagent Kits***
Output
Flow Cell Types
SP
S1
S2
S4
1 × 35 bp
N/A
N/A
N/A
280-350 Gb
2 × 50 bp
65-80 Gb
134-167 Gb
333-417 Gb
N/A
2 × 100 bp
134-167 Gb
266-333 Gb
667-833 Gb
1600-2000 Gb
2 × 150 bp
200-250 Gb
400-500 Gb
1000-1250 Gb
2400-3000 Gb
2 × 250 bp
325-400 Gb
N/A
N/A
N/A
Single read Passing Filter
650-800 M
1.3-1.6 B
3.3-4.1 B
8-10 B
Paired-end Passing Filter
1.3-1.6 B
2.6-3.2 B
6.6-8.2 B
16-20 B
*Specifications based on Illumina PhiX control library at recommended loading concentration.
Quality Scores for Reagent Kits
Run Configuration
Quality scores
(Kit type)
(SP/S1/S2/S4)
1 × 35 bp
≥ 90% of bases higher than Q30 at 1 × 35 bp
2 × 50 bp
≥ 90% of bases higher than Q30 at 2 × 50 bp
2 × 100 bp
≥ 85% of bases higher than Q30 at 2 × 100 bp
2 × 150 bp
≥ 85% of bases higher than Q30 at 2 × 150 bp
2 x 250 bp
≥ 75% of bases higher than Q30 at 2 × 250 bp
Run time estimate for Reagent Kits
Run Time*
SP
S1
S2
S4
1 × 35 bp
N/A
N/A
N/A
~14 hr
2 × 50 bp
~13 hr
~13 hr
~16 hr
N/A
2 × 100 bp
~19 hr
~19 hr
~25 hr
~36 hr
2 × 150 bp
~25 hr
~25 hr
~36 hr
~44 hr
2 x 250 bp
~38 hr
N/A
N/A
N/A
*Note: Run time includes cluster generation, sequencing, and base calling. Run times are based on running 2 flow cells of the same type; starting two different flow cells (lower and higher throughput) will impact run time.
For further information on run time estimates, check Knowledge Base articleRun time estimates for each sequencing step on Illumina sequencing platforms
See the NovaSeq 6000 System Specifications page for more information.