# Troubleshooting a Flow Cell Registration Failed (Skew) warning on the NextSeq 500/550

The following error message can occur during pre-run checks and is caused by a software communication issue and generally does not impact run performance.

The error states "Flow Cell Registration Failed (skew). No action necessary", as shown below.

![](/files/LYJ2pbhHfNXDZINDXx8h)

**Troubleshooting steps:**

* Monitor run until completion and review run specifications.
  * If run does not meet specifications, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).
* Power cycle instrument before next run.

For further information, search for Knowledge Base articles **Specifications for the NextSeq 500/550** and **How to Perform a NextSeq 500/550 Power Cycle.**

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2053), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002053%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000002053.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.