> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000002053.md).

# Troubleshooting a Flow Cell Registration Failed (Skew) warning on the NextSeq 500/550

The following warning message can occur during NextSeq 500/550 pre-run checks "Flow Cell Registration Failed (skew). No action necessary" (Figure 1).

This warning is caused by a software communication issue and generally does not impact performance.

![](/files/LYJ2pbhHfNXDZINDXx8h)

**Figure 1.** Image of the NextSeq Control Software notifications tab with the warning "Flow Cell Registration Failed (skew). No action necessary".

**Troubleshooting steps:**

1. Monitor run until completion and review run specifications. Run specifications for NextSeq 500/550 are listed here [Specifications for the NextSeq 500/550.](/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-reference_material-list/000001942.md)

* If the run does not meet specifications, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).

2. Power cycle instrument before next run. For further information, see the Knowledge Base articles [How to Power Cycle the NextSeq 500/550.](/instrumentation/general/instrumentation-general-reference_material-list/000001972.md)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2053), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001972%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000002053.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.