# Library loading concentration considerations for the iSeq 100

On the iSeq 100 the loading volume for all library types is 20 Âµl.

| Library type                       | v2 Loading Concentration (pM) | v1 Loading Concentration\* (pM) |
| ---------------------------------- | ----------------------------- | ------------------------------- |
| 100% PhiX                          | 100                           | 60                              |
| AmpliSeq for Illumina Library PLUS | 50                            | 50                              |
| Illumina DNA Prep                  | 100                           | 200                             |
| Illumina DNA Prep with Enrichment  | 75                            | 100                             |
| Nextera XT DNA                     | 150                           | 50                              |
| TruSeq DNA Nano                    | 150                           | 100                             |
| TruSeq DNA PCR-Free                | 100                           | 100                             |
| Illumina DNA PCR-Free              | 120                           | N/A                             |

For other library types, 50 pM is the recommended starting loading concentration.

**Important note:** The above recommendations are general starting points. Optimize concentrations over subsequent runs to identify a loading concentration that consistently yields data that meets run specifications.

This information can be found in the [iSeq 100 System Guide](https://support.illumina.com/downloads/iseq-100-system-guide.html).

\*v1 reagents for the iSeq 100 have been discontinued, but the information for them remains here to assist labs still finishing their v1 inventory.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2121), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002121%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-reference_material-list/000002121.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.