Library loading concentration and considerations for the iSeq 100 Instruments
The following recommendations are general starting points. Optimize concentrations over subsequent runs to identify a loading concentration that consistently yields data that meets run specifications.
The loading volume for all library types is 20 µl on the iSeq 100.
Library type
v2 Loading Concentration (pM)
100% PhiX
100
AmpliSeq for Illumina Library PLUS
50
Illumina DNA Prep
100
Illumina DNA Prep with Enrichment
75
Nextera XT DNA
150
TruSeq DNA Nano
150
TruSeq DNA PCR-Free
100
Illumina DNA PCR-Free
120
For other library types, 50 pM is the recommended starting loading concentration. Optimize this concentration over subsequent runs to identify a loading concentration that consistently yields data that meets specifications.
This information can be found in the iSeq 100 System Guide.
For any feedback or questions regarding this article (Illumina Knowledge Article #2121), contact Illumina Technical Support techsupport@illumina.com.