> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009266.md).

# Troubleshooting Pressure Charging Failures on the MiSeq i100 Series

**Background**

The MiSeq i100 Series uses pressure assisted fluidics to move reagents through the cartridge and flow cell. This requires that the wet cartridge is pressurized to a specific pressure range, and must be maintained throughout the sequencing run.

If there are issues with the reagent seals on the wet cartridge, this can cause issues with the Pre-Run Checks or during the instrument sequencing runs. If this occurs during a sequencing run, the Control Software presents the following error:

"The pressure charging failed due to air leakage detected. Contact Illumina Technical Support."

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-37793dd855511dfb413c54140760384aa9bb3e07%2Fimage1.png?alt=media)

If this issue is observed, see the following Troubleshooting Actions and contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html).

**Troubleshooting Actions**

1. Select **Eject Cartridge**, then remove the dry and wet sequencing cartridges from the instrument.
2. Power cycle the instrument to ensure all board communications are reset.
   1. See Knowledge Article [How to Power Cycle the MiSeq i100 Series](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009277.md) for more information.
3. Run the instrument Mechanical and Fluidics System Health Checks using the Test Cartridges provided with the instrument.
   1. See Knowledge Article [How to Setup the System Checks on the MiSeq i100 Series Instruments](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009206.md) for more information.
4. Once the System Health Checks are complete, select **Clear Checks** and then proceed to the Control Software home screen.
5. Use the Export Logs - Standardoption to bundle the System Health Check results.
   1. If applicable, select the checkbox next to any affected Sequencing Run(s) from the list to include the run Logs folder.
   2. See Knowledge Article [How to Export Logs on the MiSeq i100 and NovaSeq X Series Instruments](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000009265.md) for more information.
6. Once the Export Logs workflow has completed and transferred to the External Storage, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html) for further steps.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9266), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009266%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009266.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.