How to Resolve "sma_air_a/b was not present in dictionary" Errors During Pre Run Checks on the NovaSeq X Plus
Last updated
Last updated
© 2023 Illumina, Inc. All rights reserved. All trademarks are the property of Illumina, Inc. or their respective owners. Trademark information: illumina.com/company/legal.html. Privacy policy: illumina.com/company/legal/privacy.html
Background
During Pre-Run Checks, the instrument performs several tests to verify that the instrument is ready for sequencing. Any issues during this process will result in Error messages that can assist in identify the issue and resolving it. This article describes how to resolve issues with the System Thermal check that occurs during the Dry Checks section at the start of the Pre-Run Checks.
Error Message
"The given key 'sma_air_a/b' was not present in the dictionary"
Troubleshooting Actions
Important: No consumables are marked as used if this error presents during the Pre-Run Checks and consumables can be used in a follow up sequencing run.
To resolve this issue, see the following steps:
Select the Return to Start button to exit the Pre-Run Checks
Close the NovaSeq X Series Control Software:
Select the instrument icon in the bottom-middle of the Home screen to open the global navigation menu.
Select Settings.
Select Exit Software.
Restart the NovaSeq X Series Control Software:
From the Linux Operating System (OS), select the Activities menu in the top-left corner.
Select the icon for the NovaSeq X Series Control Software.
Once the Control Software has initialized, proceed with restarting run setup.