> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000003034.md).

# Storage temperature, shipping conditions and dimensions of NextSeq 1000/2000 reagents

Summary of Sequencing Consumable Shipping Condition, Storage Temperature and Packaging Dimensions.

| Consumable                                                                      | Shipping Condition | Storage Temp   | Dimensions\*                                                       |
| ------------------------------------------------------------------------------- | ------------------ | -------------- | ------------------------------------------------------------------ |
| <p>Standard SBS Reagent Cartridge<br>100-cycle, 200-cycle, or 300-cycle</p>     | Dry Ice            | -25°C to -15°C | <p>29.2 cm x 17.8 cm x 12.7 cm<br><br>(11.5 in x 7 in x 5 in)</p>  |
| <p>Standard SBS Reagent Cartridge<br>600-cycle and all XLEAP-SBS Cartridges</p> | Dry Ice            | -25°C to -15°C | <p>33.2 cm x 17.8 cm x 12.7 cm<br><br>(13.1 in × 7 in × 5 in)</p>  |
| Flow Cell                                                                       | Room Temp          | 2°C to 8°C     | <p>21.6 cm x 12.7 cm x 1.9 cm<br><br>(8.5 in x 5 in x 0.75 in)</p> |
| Resuspension Buffer (RSB) with Tween 20                                         | Room Temp          | 2°C to 8°C     | <p>4 cm x 6.6 cm x 5 cm<br><br>(1.6 in x 2.6 in x 2 in)</p>        |

\*All measurements include all packaging.

Store cartridges with the arrows pointing up, in their original box, to avoid tearing or puncturing the foil bag.

For part numbers, see [Consumable part numbers for the NextSeq 1000/2000.](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000002390)

See the [Consumables and Equipment](https://support-docs.illumina.com/IN/NextSeq10002000/Content/IN/NextSeq2000_1000/SequencingConsumables_fNS_m2000_m1000.htm) section of NextSeq 1000/2000 Product Documentation for more information.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #3034), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000003034%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000003034.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
