Troubleshooting MiSeq "Z Motor The Z stage amp Op amp might not be wired correctly" Error

Users may encounter a "Z stage/amp/Op-amp might not be wired correctly" error at any point during a MiSeq run. This issue may be caused by a communication error or by issues with the Z-stage itself.

To troubleshoot this issue :

  1. Perform the post-run wash as indicated.

  2. Power Cycle the MiSeq. Instructions for this procedure may be located in the Illumina Knowledge Article How to shut down or power cycle the MiSeqarrow-up-right.

  3. Navigate to the MiSeq System Checksarrow-up-right Menu and perform the Z-Stage test 5 times. The rest of the checks may be de-selected.

  4. If error persists, instrument fails to initialize, or one of the Z-Stage test replicates fail, Contact Illumina Technical Support.

  5. If the instrument initializes without error and all five Z-Stage tests pass, the instrument is clear for the next run.

  6. If the next run fails with the same error, contact Illumina Technical Supportarrow-up-right.