> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miniseq/instrumentation-miniseq-troubleshooting-list/000005678.md).

# Troubleshooting Windows 10 needs to be activated error on the MiniSeq

Users may notice a Windows error "**Error code: 0xc004F025 Windows 10 needs to be activated.**" on the MiniSeq. The error may appear in the bottom right corner of the instrument as "Windows 10 needs to be activated."

In the Windows Settings there is an option to manually enter a key. However, Windows 10 typically activates internally. If the MiniSeq instrument is left idle for a length of time and the password for the sbsadmin account expires, it is possible for the internal activation to be stalled.

**Troubleshooting Steps**:

1. Update the credentials for the Windows admin account (sbsadmin).
2. Open the command terminal as an administrator to manually activate Windows.
   1. Run Command prompt as administrator.
   2. Enter the command, for eg, `slmgr -rearm` then press enter key.
3. Power cycle the instrument. For further information, search for Knowledge Base article **How to Power Cycle the MiniSeq**.

If the error persists, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #5678), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000005678%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miniseq/instrumentation-miniseq-troubleshooting-list/000005678.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.