> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009382.md).

# Waste Volumes for MiSeq i100 Series Reagents

**Background**

The volume of waste produced by a MiSeq i100 Series sequencing run depends on the wet cartridge types as well as whether the instrument is set to automatically purge unused reagents when the run is complete. The flowcell type (e.g., 5M, 25M, etc.) does not affect the waste volumes.

The table below shows the expected approximate waste volumes when using the wet cartridge type specified in the table. The waste volumes depend on whether the **Purge reagent cartridge after run** checkbox is selected or not.

When the **Purge reagent cartridge** option is de-selected, the total waste volume can vary based on the total number of cycles performed in the run, with shorter runs generating reduced waste volume.

| **Flow Cell Type** | **Cycle Number** | <p><strong>Purge Enabled</strong><br><strong>Waste Volume (mL)</strong></p> | <p><strong>Purge Disabled</strong><br><strong>Waste Volume (mL)\*</strong></p> |
| ------------------ | ---------------- | --------------------------------------------------------------------------- | ------------------------------------------------------------------------------ |
| 25M or 50M         | 100              | 230                                                                         | 82-97                                                                          |
| 100M               | 100              | 385                                                                         | 106-126                                                                        |
| 5M, 25M or 50M     | 300              | 230                                                                         | 141-169                                                                        |
| 100M               | 300              | 385                                                                         | 187-225                                                                        |
| 5M, 25M or 50M     | 600              | 385                                                                         | 241-291                                                                        |
| 25M                | 1000             | 385                                                                         | 283-333                                                                        |

*\*\*\*Waste volumes will vary slightly depending on whether index cycles are performed or not.*

**Representative reference images for the respective volumes when Purge is Enabled and Disabled:**

**100 cycles Cartridge:**

-Purge Enabled (230 mL)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-ef75b094471b5545f673f962bf316896345ac89d%2Fimage1.png?alt=media)

-Purge Disabled (82-97 mL)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-e2174f122cb0b14510a5cfb003cbfb35d55fc8eb%2Fimage2.png?alt=media)

**300 cycles Cartridge:**

-Purge Enabled (230 mL)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-ef75b094471b5545f673f962bf316896345ac89d%2Fimage3.png?alt=media)

-Purge Disabled (141-169 mL)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-5c34dc270e5e85921efd48b420a0d5660529a3e1%2Fimage4.png?alt=media)

**600 cycles Cartridge:**

-Purge Enabled (385 mL)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-461770647e11795156aec746409e072eb149da65%2Fimage5.png?alt=media)

-Purge Disabled (241-291 mL)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-52dcb9293f4e93f242a5f452c479c8c1351e7bd4%2Fimage6.png?alt=media)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9382), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009382%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009382.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.