# Reagent Stability Guidelines for the MiSeq i100 Series Consumables

**Background**

The MiSeq i100 Series consumables are shipped and stored at room temperature, so there are no thawing steps required when preparing a sequencing run. However, once the packaging has been opened, the reagent kit and associated consumables should be used within the time specified below for optimal performance.

**MiSeq i100 Dry Cartridge**

Keep the Dry Cartridge in the foil packaging until it is ready to be loaded. The Dry Cartridge must be used within **4 hours** of opening the foil packaging.

**MiSeq i100 Wet Cartridge**

Keep the Wet Cartridge in the foil packaging until it is ready to be loaded, and stored vertically and in package to prevent leaks. The Wet Cartridge must be used within **4 hours** of opening the foil packaging.

**Denatured Libraries (Libraries + KLD Buffer)**

Store the denatured libraries on ice after mixing the diluted libraries with the KLD buffer. The denatured libraries are stable on ice or at 4Â° C for up to **6 hours**.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9351), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009351%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009351.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.