> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000003339.md). Agents can also search and read this documentation through its [MCP server](https://knowledge.illumina.com/~gitbook/mcp).

# Index color balancing for the NextSeq 1000/2000 system using standard SBS reagents

**Background**

When pooling libraries for sequencing on the NextSeq 1000/2000 system using standard SBS reagents, it is important to select an appropriate index combination to avoid cluster registration failures during the index read(s).

This article provides guidelines for pooling Illumina libraries on the NextSeq 1000/2000 system using standard SBS reagents. For guidance using XLEAP SBS reagents, see [Index color balancing for XLEAP-SBS reagents on the NextSeq 1000/2000 and NovaSeq X Series](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000008422.md).

**NextSeq 1000/2000 2-Channel Sequencing**

The NextSeq 1000/2000 system utilises two dyes and two images to encode data for the four nucleotides. The NextSeq 1000/2000 uses one image from the green channel and one image from the blue channel, while other two-channel systems (such as the NextSeq 500/550) use one image from the green channel and one image from the red channel.

NextSeq 1000/2000 2-channel sequencing with standard SBS kits uses two fluorescent dyes in four combinations: Blue for C, Green for T, Green and Blue for A, and no dye for G.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-13dfe951de2551f665d60128314f8ab6da938e51%2Fimage1.png?alt=media)

Figure 1: NextSeq 1000/2000 2-channel standard SBS fluorescent imaging. Clusters with intensity in green are T-bases. Clusters with intensity in blue are C bases. Clusters with intensity in green and blue are A bases. Clusters with no intensity/dark are G bases.

**Index Considerations**

* Index reads must begin with at least one base other than G in either of the first two cycles. A signal must be present in either of the first two index cycles to ensure successful sequencing. If an index read begins with two G bases, no signal intensity is generated, and cluster registration will fail.
* Select index sequences that provide signal in at least one channel, preferably both channels, for every cycle.
  * Blue channel - A or C
  * Green channel - A or T

**Index Combinations Examples**

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-a72c30951a6a546c2c6a2edc8be1d23d2ee7e4e7%2Fimage2.jpg?alt=media\&token=dc2b6ab7-343f-4eae-a4bc-0e120cf32a6e)

Ideal index combinations: contain signal in both channels for every cycle.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-c4f56a9dbad6a37085d97dfcab6d7e4b63c85776%2Fimage3.jpg?alt=media)

Acceptable index combinations: have signal in only one channel, but there is still enough signal to sequence.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-118cbc01e916d85f2703eeba2f93d3f0b5f159be%2Fimage4.jpg?alt=media)

Unacceptable index combinations: will fail registration because the index read begins with 2 base calls of G and no signal intensity is generated.

Additional index combinations can be reviewed [here](https://support.illumina.com/downloads/index-adapters-pooling-guide-1000000041074.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #3339), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000003339%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000003339.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.