> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000005229.md).

# How to set up a PhiX validation run on the NextSeq 1000/2000 sequencing platform

**Background**

* A PhiX validation run confirms proper hardware and software performance of the instrument. The [PhiX Control v3 Library](https://knowledge.illumina.com/library-preparation/general-library-prep/library-preparation-general-library-prep-reference_material-list/000001545.md) is a well-balanced genome with relatively equal representation of A, C, G, and T nucleotides. PhiX lacks an index and is not an appropriate tool for assessing Index Read performance.
* The optimal loading concentration for a PhiX validation run on the NextSeq 1000/2000 system is 650 pM for Standard chemistry and P1/P2 XLEAP Chemistry. For all P3/P4 XLEAP chemistry kits and P1/P2 600 cycle kits the optimal loading concentration is 450 pM.
* Use the following instructions to set up a PhiX validation run for the NextSeq 1000/2000 in either **Hybrid mode** or **Local mode**.
* If the instrument is connected to the Internet, set up the run in the Cloud or Hybrid mode with "Proactive and Run Monitoring" or "Proactive, Run Monitoring and Storage" (recommended for easy sharing with Illumina Support). If the instrument is not connected to the Internet or the use of BaseSpace Sequence Hub is blocked, Local or Standalone run setup mode is required.

**Hybrid Mode**

1. Log into [BaseSpace Sequence Hub](https://login.illumina.com/platform-services-manager/?rURL=https://basespace.illumina.com\&clientId=basespace\&clientVars=aHR0cHM6Ly9iYXNlc3BhY2UuaWxsdW1pbmEuY29tL2Rhc2hib2FyZA\&redirectMethod=GET#/).
2. Navigate to the Runs tab.
3. Select **New Run** on the top-right side.
4. Select **Run Planning**.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-251f339b7c9ef2c93992353b789e5093bfdc1470%2Fimage1.png?alt=media)

5. Enter a Run name.
6. Select **NextSeq 1000/2000** for Instrument Platform.
7. Select **Local Analysis**.
8. Select **Next**.

9.Select **DRAGEN BCL Convert,** ensuring that the version chosen matches the version on the NextSeq 1000/2000 instrument.

10. Select **Not Specified** for Library Prep Kit and Index Adapter Kit.
11. Select **Next**.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-793c275691c2567780e25106010802cecf58bedc%2Fimage2.png?alt=media)

12. For Index Reads, select **No Index**. PhiX does not contain index adapters so it will not successfully sequence through index reads.
13. For Read Type, Read 1 length, and Read 2 length, select the best option for the PhiX test run being performed.
14. Enter **PhiX** into the SampleID section under Sample Data.
15. Keep all default analysis settings.
16. Select **Next**.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-05a9117af256706e7da7c70b055e341cb7804e9b%2Fimage3.png?alt=media)

17. Select **Save as planned**.

**Note**: If **Local Mode** is used for run setup, use the **Export** option instead to generate a sample sheet for manual run setup.

18. When run set up is complete, and the run shows in the planned runs list in BaseSpace Sequence Hub, complete run setup on the instrument.
19. In the NextSeq 1000/2000 Control Software, select the hamburger menu in the top-left corner.
20. Select **Settings**.
21. Ensure that **Online Run Setup** is selected.
22. Verify **Proactive and Run Monitoring** is selected for BaseSpace Option.
23. Select **Save**.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-fc049a759935d35633230e4f005ad96489ddc18d%2Fimage4.png?alt=media)

24. Start run setup.
25. Sign into BaseSpace and choose appropriate Workgroup, if applicable.
26. Select the PhiX run created under the Select Run screen.
27. Select **Review**.
28. Confirm all run configurations and output location.
29. Proceed with loading the cartridge and remaining run setup.
30. Once the run is started, return to BaseSpace and generate a share link to provide to Illumina Support for run monitoring.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-6451a408a5a56b1125b402beb89e45fa6c6d4d34%2Fimage5.png?alt=media)

**Local Mode**

* It is recommended to use BaseSpace Sequence Hub to generate the sample sheet for Local Mode run setup.
* Use the **Hybrid Mode** instructions above (until step 17) to generate a sample sheet before starting run setup on the NextSeq 1000/2000.

1. After the Sample Sheet is exported, save it to a network drive connected to the instrument or a thumb drive connected to the instrument.

**Note**: It is important to keep the Sample Sheet saved on the network drive or thumb drive as saving it locally on the instrument can cause errors.

2. In the NextSeq 1000/2000 Control Software select the hamburger menu on the top-left corner.
3. Select **Settings**.
4. Ensure that **Local Run Setup** is selected.
5. Verify that **None** is selected for BaseSpace Option.
6. Select **Save**.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-7115bbb3ddd99abc6a9f4699078cce5a6f39df2a%2Fimage6.png?alt=media)

7. Start run setup.
8. On the SELECT RUN screen, select **Start with Sample Sheet** and navigate to the sample sheet, then select **Open**.
9. Select **Setup**.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-2845d9e29140de1b2ff30a5f8d1e4b722a9dbb81%2Fimage7.png?alt=media)

10. On the run review screen, confirm run configuration and output folder location.
11. Proceed with loading the cartridge and the rest of run setup.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #5229), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000005229%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000005229.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.