> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000007830.md).

# NextSeq 1000/2000 Control Software v1.5.0 installation instructions

Please follow all steps below for a successful software (SW) installation.

**Note**: If version (v) 1.2 or higher version software is currently installed, the user can upgrade to v1.5.0 Control Software directly. If v1.1 software is currently installed, the user must first upgrade to v1.2 software following the instructions found in the v1.2 software release notes.

**PREPARATION**

**These instructions are intended for the v1.5.0 Control Software upgrade only.**

1. The total installation time (not including file download times) may take up to 20 minutes.
2. Make sure that you have the password to the **ilmnadmin** account.
3. To log in to the **ilmnadmin** account:
   1. If you are already logged in as **ilmnuser** and in control software, select the control software menu, and then select **Exit Application** to access the desktop.
   2. Select the power button icon in the upper right corner and log out of **ilmnuser**.
   3. After you are on the login screen, select **ilmnadmin**, and then enter the password to log in.
   4. The control software automatically launches once you are logged in.
4. The upgrade installation files require up to **3 GB** of free space on your desktop. See file sizes below.
   1. To access your desktop, select the control software menu in the control software, and then select **Minimize Application**.
   2. Delete or transfer files off the desktop that are not needed.
   3. Empty trash to free space from the drive.
      * Select **Applications**, and then **Files** from the upper left corner of the desktop to launch the file explorer.
      * On the left panel of the file explorer, right click the **Trash** icon and select **Empty Trash**.
5. Launch Chromium to download the software upgrade installers from the Illumina Support Site.
   1. While in desktop view, open the **Applications** menu on the upper left corner. Select **Internet**, and then select **Chromium**.
      * **Note**: After launching Chromium, you may receive a prompt to unlock your login keyring. You can safely ignore and cancel out of the prompt.
   2. Use Chromium to browse to the **NextSeq 1000 and 2000 Software Downloads** page on the Illumina Support Site.
6. Download the following software upgrade installers from the Illumina Support Site and save them to the desktop. **Do not extract these files; leave them as .tar.gz.**
   1. For offline customers:
      * v1.5.0 Software (SW) Suite Installer (1 GB)
7. After the files have finished downloading to the desktop, maximize the control software by selecting it from the toolbar.
8. To begin installation, follow the the **Offline Installation** steps.

**ONLINE INSTALLATION**

**Follow the “Preparation” steps outlined above before continuing.**

1. Install SW suite.
   1. Select the software update alert in the control software notification panel and follow the prompts to upgrade the software suite.
   2. When the installation process completes, you will see a message that confirms installation. After closing the prompt, the control software automatically relaunches.
   3. Follow the prompt for the firmware update.
      * **Note**: During this process, the screen will go blank. Wait for the image to return and the process to complete.
   4. Follow the prompt to power cycle the instrument.
      * **Note**: Make sure to toggle the power switch at the back of the instrument to complete the power cycle after the instrument shuts down.
   5. Power on the instrument and log in as **ilmnadmin**. The control software automatically launches.
      * **Note**: An alert appears indicating that **DRAGEN** workflows need to be updated. Proceed through the next steps to resolve the alert.
   6. Confirm that the version is updated to v1.5.0 in the About screen in the control software **menu**.
   7. Proceed to the next step to install **DRAGEN** workflows.

**OFFLINE INSTALLATION**

**Follow the “Preparation” steps outlined above before continuing.**

1. Install SW suite.
   1. Launch control software
      * Open control software **menu**.
      * Select Software **Update**.
      * Select **Choose**… under **Browse for Software Update**.
      * Navigate to the desktop and select the installer
      * Select **Update Now**.
   2. Follow the password and other prompts.
   3. When the installation process completes, you will see a message that confirms installation. After closing the prompt, the control software automatically relaunches.
   4. Follow the prompt for the firmware update.
      * **Note**: During this process, the screen will go blank. Wait for the image to return and the process to complete. If the image does not return after 5 minutes, proceed to the next step to manually toggle the power switch.
   5. Follow the prompt to power cycle the instrument.
      * **Note**: Make sure to toggle the power switch at the back of the instrument to complete the power cycle after the instrument shuts down.
   6. Confirm that version is updated in the **About** screen in the control software menu.
2. Final steps.
   1. After installation completes, delete all the downloaded installation files from the desktop and empty the trash to save space.
   2. Log back into **ilmnuser** to begin sequencing.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7830), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007830%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000007830.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.