Troubleshooting a CavroSyringePump Error on the NextSeq 500/550

This error is often due to a communication issue on the instrument; however, it can also result from an issue with the fluidics system on the NextSeq 500/550. * "An error occurred while sequencing. Will now proceed to determining whether to perform a post-run wash. Originally caused by: Error in Fluidics.FlowCellFluidics1.LanePump.IXPassThroughCavroSyringePump2. Auto syringe overload recovery for dispense is not enabled."

**Troubleshooting steps:**1. Upon run completion, power cycle the instrument and verify the initialization completes without error. 1. From the instrument home screen, select Manage Instrument. 2. Select System Shut down Options. 3. Select Shut Down (this will shut down software and turn off instrument power). 4. Wait two minutes before turning on the instrument from the button in front. 5. Wait for software initialization back to the Home Screen.

  1. Should the system initialize without error, perform a System Check* to verify the function of the fluidics system.

  1. If the checks pass, perform a Manual Post-Run Wash before setting up the next run.

*For further information, search for Knowledge Base articles How to perform a NextSeq 500/550 system check andHow to perform washes on the NextSeq 500/550.

For any feedback or questions regarding this article (Illumina Knowledge Article #1380), contact Illumina Technical Support techsupport@illumina.com.

Last updated

© 2023 Illumina, Inc. All rights reserved. All trademarks are the property of Illumina, Inc. or their respective owners. Trademark information: illumina.com/company/legal.html. Privacy policy: illumina.com/company/legal/privacy.html