> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001380.md).

# Troubleshooting a CavroSyringePump Error on the NextSeq 500/550

**Background**

This error is often due to a communication issue on the instrument. However, it can also result from an issue with the fluidics system on the NextSeq 500/550.

* "An error occurred while sequencing. Will now proceed to determining whether to perform a post-run wash. Originally caused by: Error in Fluidics.FlowCellFluidics1.LanePump.IXPassThroughCavroSyringePump2. Auto syringe overload recovery for dispense is not enabled."

**Troubleshooting steps:**

1. Upon run completion, power cycle the instrument and verify the initialization completes without errors.
   1. From the instrument home screen, select **Manage Instrument.**
   2. Select **System Shut down Options.**
   3. Select **Shut Down** (this shuts down the software and turns off instrument power).
   4. Wait two minutes before turning on the instrument from the button in front.
   5. Wait for software initialization back to the Home Screen.
   * If the system fails to initialize, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for additional assistance.
2. Should the system initialize without error, perform a System Check to verify the function of the fluidics system. Reference article [How to perform a NextSeq 500/550 system check](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001883.md).
   * If the system check fails the Volume tests, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for additional assistance.
3. If the checks pass, perform a Manual Post-Run Wash before setting up the next run. Reference article [How to perform washes on the NextSeq 500/550](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-reference_material-list/000003076.md).
   * If additional errors occur during the wash, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for additional assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1380), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001380%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001380.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.